IS 12-TSP IGA 12-TSP IMPAC-Transfer-Standard-Pyrometer

Größe: px
Ab Seite anzeigen:

Download "IS 12-TSP IGA 12-TSP IMPAC-Transfer-Standard-Pyrometer"

Transkript

1 IS 12-TSP IGA 12-TSP IMPAC-Transfer-Standard-Pyrometer Operation Manual Betriebsanleitung Proven Quality

2 IMPAC pyrometers IS 12-TSP IGA 12-TSP LumaSense Technologies All rights reserved. LumaSense Technologies 3033 Scott Blvd. Santa Clara, CA Tel.: Fax: Internet: LumaSense Technologies 16 Thornton Road Oakland, NJ Tel.: Fax: Internet: Please call for Factory Repair and Return There are no user-serviceable components in the instrument Disassembly of the instrument is not allowed, and the warranty is VOID if the instrument is disassembled, tampered with, altered or otherwise damaged, without prior written consent from LumaSense Technologies or if considered by LumaSense Technologies to be abused or used in abnormal conditions. No adjustments may be made to the targeting laser. It is fixed at the factory. No adjustments may be made to the targeting laser s power level. General Information about the user manual Congratulations on choosing the high quality and highly efficient IMPAC pyrometer. Please read this manual carefully, step by step, including all notes to security, operation and maintenance before installing the pyrometer. For installation and operation of the instrument this manual is an important source of information and work of reference. To avoid handling errors keep this manual in a location where you always have access to. When operating the instrument it is necessary to follow the general safety instructions (see section 3, Safety). Additionally to this manual the manuals of the components used are valid. All notes especially safety notes are to be considered. Limit of liability and warranty All general information and notes for handling, maintenance and cleaning of this instrument are offered according to the best of our knowledge and experience. LumaSense Technologies is not liable for any damages that arise from the use of any examples or processes mentioned in this manual or in case the content of this document should be incomplete or incorrect. LumaSense Technologies reserves the right to revise this document and to make changes from time to time in the content hereof without obligation to notify any person or persons of such revisions or changes. All series 12 instruments from LumaSense Technologies have a warranty of two years from the invoice date. This warranty covers manufacturing defects and faults which arise during operation only if they are the result of defects caused by LumaSense Technologies. The Windows compatible software was thoroughly tested on a wide range of Windows operating systems and in several world languages. Nevertheless, there is always a possibility that a Windows or PC configuration or some other unforeseen condition exists that would cause the software not to run smoothly. The manufacturer assumes no responsibility or liability and will not guarantee the performance of the software. Liability regarding any direct, or indirect damage caused by this software is excluded. Copyright All copyrights reserved. This document may not be copied or published, in part or completely, without the prior written permission of LumaSense Technologies GmbH. Contraventions are liable to prosecution and compensation. All rights reserved. 2

3 IMPAC pyrometers IS 12-TSP IGA 12-TSP Contents General... 2 Information about the user manual... 2 Limit of liability and warranty... 2 Copyright... 2 Legend... 5 Terminology... 5 Disposal / decommissioning Technical data Dimensions Overview Appropriate use Scope of delivery Safety General Laser targeting light Electrical connection Electrical Installation Pin assignment for the connector on the back of the pyrometer for main connection cable for additional limit contacts cable Connecting the pyrometer to a PC Connecting to RS232 interface Connecting to RS485 interface Connection of additional analyzing devices Mechanical installation Aligning of the pyrometer to the aperture Special aligning notes Instrument settings Sighting Thru-lens view finder Laser targeting light Optics Adjusting the required measuring distance Adjusting the measuring distance with help of the table Other measuring distances Instrument settings Key panel operation Functions and setting of parameters Factory settings

4 IMPAC pyrometers IS 12-TSP IGA 12-TSP 9 Parameters Emi (emissivity ) t 90 / s (exposure time) t CL / s (clear times of the maximum value storage) FROM / TO (beginning and end of sub range) / 4 ma (analog output) SETP (limit contacts) C / F (temperature display) RS (digital interface) ADR (address) kbaud (baud rate) ma (test current output) Wait time (only available via interface commands, see 15, Data format UPP ) Switch contact Settings via interface and software Installation Program start The start menu Beginning Number of devices Basic settings Measurement (color bar) Measurement (online trend) Output listing (analyzing) Output.TXT file (analyzing) Output trend (analyzing) PC sampling rate (time interval between two measurements) Spot size calculator Transport, packaging, storage Maintenance Safety Service Changing of optics Trouble shooting Data format UPP (Universal Pyrometer Protocol) Reference numbers Reference numbers instruments Reference numbers accessories Index

5 IMPAC pyrometers IS 12-TSP IGA 12-TSP Legend MB Note: The note symbol indicates tips and useful information in this manual. All notes should be read with regard to an effective operation of the instrument. Security note laser beam Indicates to the danger of a built-in laser targeting light. Shortcut for Temperature range (in German: Messbereich) Terminology The used terminology corresponds to the VDI- / VDE-directives 3511, page 4. Disposal / decommissioning Inoperable IMPAC pyrometers have to be disposed corresponding to the local regulations of electro or electronic material. 1 Technical data Temperature ranges: IS 12-TSP: 600 to 2520 C IGA 12-TSP: 200 to 1020 C 600 to 3000 C 250 to 1400 C 850 to 2520 C Spectral range: IS 12-TSP: 0.94 µm (at temperature range 600 to 2520 C / 3000 C) 0.65 µm (at temperature range 850 to 2520 C) IGA 12-TSP: 1.57 µm Accuracy 1 ( = 1, t 90 = 1 s, T amb. = 23 C) Repeatability: ( = 1, t 90 = 1 s, T amb. = 23 C) Below 1500 C: 0.15% of measured value in C + 1 C Above 1500 C: 0.25% of measured value in C Above 2700 C: 0.35% of measured value in C 1 C Resolution: Up to 1000 C: 0.1 C at interface and display Above 1000 C: 0.1 C at display, 0.01 C at interface < 0.025% of temperature range at the analog output Power supply: Power consumption: Analog output: Load: Test current output: Digital interface: Display: Isolation: Sub range: Signal processing: Parameters: 24 V DC (15 to 40 V DC) or 24 V AC (12 to 30 V AC), 48 to 62Hz max. 14 W 0 to 20 ma or 4 to 20 ma switchable 0 to 500 Ohm 10 ma, fixed Switchable at the instrument: RS232 or RS485 addressable (half duplex), baud rate 2.4 up to kbd Built-in 5 digit LED display, height 13 mm, additional function LED s Power supply, digital interface, analog output are galvanically isolated against each other and housing Any range adjustable within the temperature range minimum span 51 C Photoelectric current, digitized Adjustable at the instrument or via serial interface: emissivity ; response time t 90 ; clear times t CL of the maximum value storage; temperature sub range; analog output 0 to 20 or 4 to 20 ma; switch points for limit contacts; C or F; interface RS232 or RS485; 1 Note: A surrounding temperature of 10 to 43 C is permissible for measured temperatures between 10% and 100% of the temperature range for the stated measurement uncertainty. The measurement uncertainty increases by 1 C for surrounding temperatures of between 0 to 10 C or 43 to 60 C. For temperature ranges of up to 5% of the measuring range an additional temperature drift of 0.2 C/ C (23 C) should be added. The temperature drift is 0.05 C/ C (23 C) for measured temperatures of 5% to 10% of the measuring range. 5

6 IMPAC pyrometers IS 12-TSP IGA 12-TSP address; baud rate; test current output Additionally (only via interface adjustable): key lock, wait time, recalibration (with special software) Emissivity : to in 1/1000 steps Exposure time t 90 : 1 ms (with dynamical adaptation at low signal levels); factory setting 1 s, adjustable to 0.01 s; 0.05 s; 0.25 s; 1 s; 3 s; 10 s Maximum value storage: Built-in single or double storage. Clearing with adjusted time t clear (off; 0.01 s; 0.05 s; 0.25 s; 1 s; 5 s; 25 s), extern, via interface or automatically with the next measuring object Limit switches: 2 relay outputs (switch-over relay contacts), switch power max. 30 W (I max : 1 A, U max : 60 V DC) Rise and fall times: exposure time of the pyrometer + 2 ms Control panel: 4 keys, operate with tip of ball-point pen Protection class: IP65 (DIN ) Ambient temperature: 0 to 70 C at housing Storage temperature: -20 to 70 C Relative humidity: Non condensing conditions Weight: 2.2 kg Sighting: built-in parallax free thru-lens view finder, optionally with an additional built-in laser targeting light (max. power level < 1 mw, = nm, CDRH class II) CE-label: According to EU directives about electromagnetic immunity 1.1 Dimensions 6

7 IMPAC pyrometers IS 12-TSP IGA 12-TSP 2 Overview Display C or F as well as active limit switches 2 LED display for meas. temperature or parameters 3 Parameter indicator 4 Parallax free thru-lens view finder 5 Setting keys (operate with tipp of ball-point pen 6 Connection cable 7 Laser warning label (only instruments with laser targeting light) 8 Focusable optics 9 Label with optics information 10 Mounting holes for accessories 12 Type label (bottom side of the instrument) 2.1 Appropriate use IS 12-TSP and IGA 12-TSP are extremely precise and longer-term stable transfer standard pyrometers which can be used for the checking of calibration sources in the respective pyrometer spectral range. 2.2 Scope of delivery Instrument with selectable optics, PC measurement and evaluation software InfraWin, works certificate with 3 selected measuring points, user manual. Note: The connection cable is not included with the instrument and has to be ordered separately (see 15, Reference numbers). 3 Safety This section offers an overview about important safety aspects. Additionally in the several sections there are concrete safety aspects to avert danger. These aspects are indicated with symbols. Labels and markings at the instrument have to be noticed and keep in a permanent readable condition. 3.1 General Each person working with the pyrometer must have read and understood the user manual before operation. Also this has to be done if the person already used similar instruments or was already trained by the manufacturer. The pyrometer has only to be used for the purpose described in the manual. It is recommended to use only accessories offered by the manufacturer. 3.2 Laser targeting light For easy alignment to the measuring object the pyrometers can be equipped with a laser targeting light. This is a visible red light with a wavelength between 630 and 680 nm and a maximum power of 1 mw. The laser is classified as product of laser class II. 7

8 IMPAC pyrometers IS 12-TSP IGA 12-TSP Warning: To reduce the risk of injury to the eyes, do not look directly into the targeting laser and do not point the targeting laser into anyone's eyes. The instrument is equipped with a class II laser that emits radiation. Safety regulations: Never look directly into the laser beam. The beam and spot can be watched safely from side. Make sure that the beam will not be reflected into eyes of people by mirrors or shiny surfaces. 3.3 Electrical connection Follow common safety regulations for mains voltage (230 or 115 V AC) connecting additional devices operating with this mains voltage (e.g. transformers). Touching mains voltage can be fatal. An incorrect connection and mounting can cause serious health or material damages. Only qualified specialists are allowed to connect such devices to the mains voltage. 4 Electrical Installation The IS 12-TSP or IGA 12-TSP are powered by a voltage of either 24 V (optional 15 to 40 V DC or 12 to 30 V AC, 48 to 62 Hz. As the instrument is operating with heated and thermostatically controlled sensors, it needs a warm-up of up to 30 minutes to achieve the maximum accuracy. But directly after connecting to the power supply the instrument already has a very good accuracy due to the standard temperature compensation. To switch off the instrument, unplug the connector or interrupt the power supply. To meet the electromagnetic requirements, a shielded connecting cable must be used. The shield of the connecting cable has to be connected only on the pyrometer side to avoid ground loops. LumaSense offers connecting cables, they are not part of standard scope of delivery. The main connecting cable has wires for power supply, interface, analog output, external laser switch and external clear of maximum value storage via contact (see 15, Reference numbers) and 12 pin angle connector. The cable includes a short RS232 adapter cable with a 9 pin SUB-D connector for direct PC communication. This adapter is not used in combination with RS485 interface. connector connector For use of the limit contacts the separate additional cable has to be for main for additional cable used. connection cable for limit switches 4.1 Pin assignment for the connector on the back of the pyrometer for main connection cable Pin Color Indication K white + 24 V power supply (or 24 V AC) ( V) A brown 0 V power supply L green + I outp. analog output B yellow I outp. analog output H gray external switch for targeting light (bridge to K) J pink external clearing of maximum value storage (bridge to K) * ) or output for switch contact (see 9.13) G red DGND (Ground for interface) F black RxD (RS232) or B1 (RS485) C violet TxD (RS232) or A1 (RS485) D gray/pink B2 (RS485) (bridge to F) E red/blue A2 (RS485) (bridge to C) M orange Screen only for cable extension don t connect at the switchboard * ) for setting of clear time to "extern" (see 9.3 clear time for maximum value storage) C D E B L F A G M K J H Pin assignment of left connector (side of male inserts) 8

9 IMPAC pyrometers IS 12-TSP IGA 12-TSP for additional limit contacts cable The instrument is equipped with two independent relay limit switches. These are two separate switch-over relay contacts, changing its state if the adjusted temperatures are exceeded. Pin Color Indication 7 pink 5 white 1 yellow 3 grey 6 green 4 brown 2 blue 8 red Limit contact S1 (drawing shows status without power or with exceeded limits) Limit contact S2 (drawing shows status without power or with exceeded limits) Pin assignment of right connector (side of male inserts) The drawing of the limit contacts indicates their switch status without power. The limit contacts switch after supplying the pyrometer with power. If the temperature exceeds the adjusted limit value, the corresponding limit contact switches back in the first position and is indicated by the corresponding LED on the pyrometer (see 8.2 under Limit contacts). Any temperature value within the range of the pyrometer is adjustable. The setting of the limits can be done directly at the pyrometer (see 9.6, SETP (limit contacts)) or via PC and software InfraWin (see 10.6, Basic settings or 10.7, Measurement color bar. The switch time of the relay contacts is 2 ms plus the response time of the pyrometer. 4.2 Connecting the pyrometer to a PC The pyrometers are equipped with a serial interface RS232 or RS485 (switchable at the pyrometer) for connection to a processor / PC. Standard on a PC is the RS232 interface. At this interface one pyrometer can be connected if the interface is set to RS232. Only short distances can be transmitted with RS232 and electromagnetic interferences can affect the transmission. With RS485 the transmission is to a large extend free of problems, long transmission distances can be realized and several pyrometers can be connected in a bus system. If RS485 is not available at the PC, it can be realized with an external converter which converts the RS485 in RS232 for a standard connection to a PC. When using a converter RS485 RS232 take care, that the converter is fast enough to receive the pyrometer s answer to an instruction of the master. Most of the commonly used converters are too slow for fast measuring equipment. So it is recommended to use the LumaSense-converter (order no ). With a slow RS485 connection it is also possible to set a wait time at the pyrometer which delay the response of a command to the pyrometer (see 9.12, Wait time tw) Connecting to RS232 interface The transmission rate (in baud) of the serial interface is dependent on the length of the cable. Values between 2400 and Bd may be set. The baud rate has to be reduced by 50% when the transmission distance is doubled (see also 9.10, Baud rate (kbaud)). Typical cable length for RS232 at Bd is 7 m. C D E B F L J A G M K H Pyrometer s side (soldering side) TxD (violet) RxD (black) DGND (red) Cable PC s side (soldering side) 9

10 IMPAC pyrometers IS 12-TSP IGA 12-TSP Connecting to RS485 interface Half-duplex mode: A1 and A2 as well as B1 and B2 are bridged in the 12-pin round connector of the connecting cable, to prevent reflections due to long stubs. It also safeguards against the interruption of the RS485 Bus system should a connecting plug be pulled out. The master labels mark the connections on the RS485 converter. The transmission rate of the serial interface in Baud (Bd) is dependent on the length of the cable. Values between 2400 and 115 kbd may be set. The baud rate is reduced by 50% when the transmission distance is doubled (see 9.10, Baud rate (kbaud)). Typical cable length for Bd is 2 km. Master A B S A1 B2 A2 B1 C D E B L F A DGND G M K J H Pyrometer 1 e.g. address 00 A1 B2 A2 B1 C D E B L F A DGND G M K J H Pyrometer 2 e.g. address 01 Terminator 120 Ohm A1 B2 A2 B1 C D E B L F A DGND G M K J H Pyrometer 32 e.g. address Connection of additional analyzing devices Additional analyzing instruments, for example a LED digital display instrument only needs to be connected to a power supply and the analog outputs from the pyrometer. Another Instruments like a controller or printer can be connected to the display in series as shown above (total load of resistance max. 500 Ohm). white 230 V ~ brown green 24 V DC C Power supply LED digital display Controller yellow Writer 5 Mechanical installation For correct temperature determination of a calibration source, the pyrometer must be aligned centrally and along the optical axis of the respective radiation source. For reasons of different sizes and construction types of calibration sources, the installation brackets from the LumaSense accessories are suitable to only a limited extent for these wide-ranging requirements. You must therefore purchase or build a suitable alignment aid yourself. The pyrometer is fitted with M5 diameter holes on the front face and bottom for fixing as required. Installation note: Strong incidence of daylight or lamp light into the view finder can affect the measurement. In this case the ocular has to be covered. 5.1 Aligning of the pyrometer to the aperture Notes: The information contained in VDI/VDE (German Electrical Engineering Association) Directive 3511, Sheet 4 and supplementary sheet 4.4 must be observed when calibrated pyrometers and when correctly carrying out temperature measurements. 10

11 IMPAC pyrometers IS 12-TSP IGA 12-TSP The following points must also be observed in the special case of the TSP device: Special aligning notes The size of the spot size is determined first and foremost by the distance of the pyrometer from the radiator in addition to the optical equipment fitted to the pyrometer (see measure field table in 7, Optics). The pyrometer s spot size should be selected so that the temperature is only received from the floor of the radiator opening, since the emissivity is normally high and homogenous in this location but normally declines along the wall and towards the opening. The pyrometer should not be positioned so close to the radiator that it is heated by this to above its maximum permissible ambient temperature. If the selected pyrometer measured field is very small, the temperature display can change due to possible non-homogenous radiator surfaces when aligning the pyrometer (this is especially noticed on radiators of not particularly high quality). In typical cases, the optical focus point is adjusted at the height of the radiator front. To achieve a smaller or larger measured field, the focus point can be adjusted in front or behind this location. Course of radiation Radiator front Measured distance Optics Notes: IMPAC transfer standard pyrometers are fitted with an extremely good optical system which only receives the minimum radiation quantities outside the measured beam. For this reason it is not necessary to put an additional aperture in the course of the beam during measurement which limits the received radiation. This should be taken into consideration especially for comparative measurements with other pyrometer types since a high proportion of unwanted additional radiation can be received under certain conditions depending on the optical system fitted if no aperture is fitted during measurement Instrument settings The pyrometer has been set to an emissivity of 1 ex works. All calibration sources which have been calibrated with an effective emissivity of 1 will be checked with this setting. In case that the calibration source has to be checked with its true emissivity, the corresponding value can be set at the pyrometer (see 8, Instrument settings). The exposure time is set to 1 sec ex works. This is the standard setting for TS pyrometers which is achieved with maximum precision at minimum noise values. Changing this setting is only necessary if the pyrometer is to be used for other measuring tasks. 6 Sighting For exact aiming to the object the pyrometers are equipped with a thru-lens view finder and an additional laser targeting light (exception: the IS 12-TSP with temperature range 850 to 2520 C has only a view finder). 6.1 Thru-lens view finder In the optimized thru-lens view finder a circle marks the exact position and size of the measuring spot. The focusable optics is correctly adjusted to the required distance, if the measuring object is shown as a sharp image in the view finder (adjusting the measuring distance see 7, Optics). Pyrometers with temperature ranges exceeding 1500 C are equipped with an adjustable eye protection filter. Turning the ocular changes the filter from bright to dark. Eye protection filter 11

12 IMPAC pyrometers IS 12-TSP IGA 12-TSP 6.2 Laser targeting light The laser point marks the center of the spot, not the exact size. On the focused measuring distance the laser has its smallest spot size and is illustrated exactly (adjusting the measuring distance see 7, Optics). The laser targeting light can be used without affecting the measurement. 7 Optics The pyrometers are equipped ex works with a focusable optics. With this focusable optics the best possible distance ratio can always be adjusted for each distance within the optics limits (the smallest and largest value of each optics for the measuring distance in the table). Optics 1 Optics 2 Optics 3 Focusable optics IS 12-TSP Measuring distance a [mm] Spot size M 98 [mm] Focusable optics IGA 12-TSP Spot size M 98 [mm] Objective length S [mm] all temperature ranges C C Aperture D [mm] * ) * ) The aperture is the effective lens diameter. It is depending on the objective length. The biggest aperture value (17 mm) belongs to the fully extended objective (S = 30 mm), the smallest aperture value (13.5 mm) for objective length S = 0. Intermediate values have to be interpolated (e.g. calculating the spot size if the measuring distance differs from the adjusted distance). The spot size calculator included in the InfraWin software is the easiest way to calculate the spot size if the measuring distance differs from a fixed adjusted (see 10.13, Spot size calculator). 7.1 Adjusting the required measuring distance The required measuring distance must be adjusted to achieve the spot sized mentioned in the tables above. For releasing the optics has to be turned anticlockwise. Then it can be pushed or pulled to find the correct measuring distance (see 6, Sighting). For fixing the optics has to be turned clockwise. Push / pull fix release Note: The measuring object has to be bigger or at least as big as the spot size of the pyrometer Adjusting the measuring distance with help of the table The table mentions the minimum and maximum measuring distance for each of the 3 focusable optics. This corresponds to the longest or the shortest objective length. As an example a further value between max. and min. is shown. The objective length S can be measured with a caliper Other measuring distances For all other measuring distances the optics will be adjusted with help of the view finder or the laser targeting light (see 6, Sighting). S (Objective length) 12

13 IMPAC pyrometers IS 12-TSP IGA 12-TSP 8 Instrument settings All instrument settings can be done directly at the instrument. To activate the adjusting keys, a tip of ball pen must be used. This avoids changing parameters by mistake. 8.1 Key panel operation Temperature- or parameter display Indication of limit switches Indication C or F Parameter indicator Setting keys: Targeting light on / off 1) Parameter selection 2) Change of parameters: 3) Confirmation of parameter Emissivity Exposure time (t 90 /s) Clear time of max. value storage (t CL /s) Beginning of sub range (FROM) End of sub range (TO) Analog output 0 or ma (0/4 ma) Setting of limit switches (SETP) Temperature in C or F ( C/ F) Interface RS232 or RS 485 (RS) Address (ADR) Baud rate (kbaud) Test current output (10 ma) 8.2 Functions and setting of parameters Temperature or parameter display: Targeting light on / off: Setting keys: 1) PAR: Limit contacts: Scale C or F: 2) : 3) ENT: Parameter indicator: In the measuring mode the display shows the actual temperature reading. After pushing the parameter key the display indicates the actual value of one selected parameter. Special indications: 8888 = measurement exceeds the end of sub range Pushing the PL key in measuring mode the laser targeting light is switched on. The IS 12 measures continuously also with switched on targeting light and the display shows the actual temperature reading. The laser targeting light is switched off automatically after approx. 2 min or after a further push of the PL key. With the PAR button all available parameters are displayed in the following description. Pushing the button again changes the display to the next parameter, after the last parameter it changes to the actual temperature reading. With the arrow keys and all parameter settings can be displayed. Pushing the button longer changes the settings in fast mode (the keys and ENT are blocked if the keyboard is locked via the interface commands or/and during a measurement in combination with a PC and InfraWin). If a parameter is changed with the arrow keys the new value must be confirmed by pushing the ENT key. If it is not confirmed with ENT the instrument is still operating with the previous parameter value. If no key is pressed for approx. 30 s the display changes to the temperature indication. Two limit switches can be set. If the measuring temperature exceeds the adjusted limit contact temperature, the LED S1 or S2 displays the switch status. The LED indicates the temperature scale in C or F. LED s indicate which pyrometer parameter is selected for reading or changing. 13

14 IMPAC pyrometers IS 12-TSP IGA 12-TSP 8.3 Factory settings Emissivity () = 1,000 Exposure time (t 90 / s) = 1 s Clear time (t CL ) = 0,00 (= OFF) Sub range (from / to) same as temperature range Analog output (0 / 4 ma) = ma Limit contacts = end of temperature range Hysteresis = ±2 C Temperature display ( C / F) = C Interface (RS) = RS232 Address (ADR) = 00 Baud rate (kbaud) = 19,2 kbd Test current output (10 ma) = off 9 Parameters 9.1 Emi (emissivity ) For a correct measurement it is necessary to adjust the emissivity. This emissivity is the relationship between the emission of a real object and the emission of a black body radiation source (this is an object which absorbs all incoming rays and has an emissivity of 1 = 100%) at the same temperature. Different materials have different emissivities ranging between 0% and 100% (settings at the pyrometer between 10 and 100%). Additionally the emissivity is depending on the surface condition of the material, the spectral range of the pyrometer and the measuring temperature. The emissivity setting of the pyrometer has to be adjusted accordingly. Typical emissivity values of various common materials for the two spectral ranges of the instruments are listed below. The tolerance of the emissivity values for each material is mainly dependent on the surface conditions. Rough surfaces have higher emissivities. Measuring object Emissivity Emissivity Emissivity (at 0.65 µm) (at 0.94 µm) (at 1.57 µm) Black body furnace Steel heavily scaled to 0.9 Steel rolling skin to 0.88 Steel, molten to 0.25 Slag to 0.85 Aluminum, bright Chromium, bright 0.32 to to to 0.3 Brass oxidized (tarnished) 0.7 to to to 0.7 Bronze, bright Copper, oxidized to 0.85 Copper, bright Zinc to 0.55 Nickel to 0.2 Gold, bright Silver, bright Porcelain glazed Porcelain rough 0.8 to to to 0.9 Graphite 0.82 to to to 0.9 Fireclay 0.45 to to to 0.6 Stoneware, glazed 0.86 to to to 0.9 Brick 0.85 to to to 0.9 Soot t 90 / s (exposure time) The exposure time is the time interval when the measured temperature has to be present after an abrupt change so that the output value of the pyrometer reaches a given measurement value. The time taken is to reach 90% of the recorded temperature difference. In the 0.00 position (Software InfraWin: min ), the device operates using its time constant. The dynamic exposition time adjustment prolongs the exposure time at the lower range limit. Settings: 0.00 s 0.01 s 0.05 s s 14

15 IMPAC pyrometers IS 12-TSP IGA 12-TSP 9.3 t CL / s (clear times of the maximum value storage) If the maximum value storage is switched on always the highest last temperature value will be displayed and stored. The storage has to be cleared at regular intervals for exchanging by a new and actual value. This feature is particularly useful when fluctuating object temperatures cause the display or the analog outputs to change too rapidly, or the pyrometer is not constantly viewing an object to be measured. In addition, it may also be beneficial to periodically delete and reset the stored maximum values. The following settings are possible: Settings: 0.00 s (= OFF) 0.01 s. 25 s extern auto Hold OFF: At clear time OFF the max. value storage is switched off and only momentary values are measured s: If any clear time between 0.01 s and 25 s is set, the maximum value is estimated and held in double storage mode. After the entered time the storage will be deleted. extern: The external clearing can be activated and used within an own software (see section 14, Data format UPP ) or via an external contact (for connection see 4.1 Connector pin assignment on the back side of the pyrometer). In this case, the storage operates only in single storage, because only a single deletion mechanism is used. auto: The auto mode is used for discontinuous measuring tasks. For example objects are transported on a conveyer belt and pass the measuring beam of the pyrometer only for a few seconds. Here the maximum value for each object has to be indicated. In this mode the maximum value is stored until a new hot object appears in the measuring beam. The temperature which has to be recognized as hot is defined by the low limit of the adjusted sub range. The stored maximum value will be deleted when the temperature of the new hot object exceeds the low limit from of the sub range by 1% or at least 2 C. If a lower limit is not entered, the maximum value storage will be deleted whenever the lower level of the full measuring range has been exceeded. Hold: The function hold enables to freeze the current temperature reading at any moment. For this an external push button or switch has to be connected (see connector pin J in 4.1) which holds the temperature reading as long as the contacts are closed. Operation note: dependent on the settings the maximum value storage either works in single storage mode or in double storage mode: Single storage: the single storage is used when you want to reset the stored value using an external impulse via one contact closure from an external relay (i.e. between two measured objects). The relay contact is connected directly to the pyrometer between pins J and K. This mode allows a new value to be established, after each impulse from the reset signal. double storage: when entering the reset intervals via push buttons or PC interface the double storage is automatically selected. This mode utilizes two memories in which the highest measured value is held and is deleted alternately in the time interval set (clear time). The other memory retains the maximum value throughout the next time interval. The disadvantages of fluctuations in the display with the clock frequency are thereby eliminated. Note: The maximum value storage follows the function of adjustment of exposition time. This results in: clear time the adjusted response time is useless clear times must be at least 3 times longer than the response time only maxima with full maximum value can be recorded, which appear at least 3 times longer than the response time. 9.4 FROM / TO (beginning and end of sub range) You have the opportunity to choose a sub range (minimum 51 C) within the basic measuring range of the pyrometer. This sub range corresponds to the analog output. FROM describes the beginning of this measuring range, TO the end of the range. Additionally with the setting of a sub range it is possible to fulfill the requirements of the auto clear mode of the maximum value storage (see above). 15

16 IMPAC pyrometers IS 12-TSP IGA 12-TSP / 4 ma (analog output) The analog output has to be selected according to the signal input of the connected instrument (controller, PLC, etc.). Settings: ma ma 9.6 SETP (limit contacts) The instrument is equipped with two independent relay limit switches. These are two separate switch-over relay contacts, adjustable to any temperature of the pyrometer s temperature range. The switch-status is indicated in the display with the LED s S1 and S2. When the measuring temperature exceeds the adjusted limit contact temperature the relay switches and the LED indication is on. 9.7 C / F (temperature display) The temperature can be displayed in C or F. 9.8 RS (digital interface) Selection between RS232 and RS485. (This adjustment can only be done on the pyrometer). Settings: C F Settings: RS232 RS ADR (address) Settings: For the connecting of several pyrometers with RS485 with one serial interface it is necessary to give each instrument an individual address for communication. First it is necessary to connect each single instrument to give it an address. After that all instruments can be connected and addressed individually. If parameters may be changed simultaneously on all pyrometers, the global Address 98 can be used. This allows you to program all pyrometers at the same time, regardless of the addresses that have already been assigned. If the address of a pyrometer is unknown, it is possible to communicate with it using the global Address 99 (connect only one pyrometer) kbaud (baud rate) The transmission rate of the serial interface in Baud (Bd) is dependent on the length of the cable. A standard cable length with RS232 for Bd is 7 m, with RS485 2 km. The baud rate is reduced by 50% if the transmission distance is doubled. Settings: 2.4 kbd kbd ma (test current output) The test function activates a test current of 10 ma on the analog output, independently if the instrument is adjusted to 0 to 20 ma or 4 to 20 ma. This enables to test external indicators for correct scaling. The reading of the internal indicator should correspond to the reading of the external indicator. If both displays show different readings the external indicator has a wrong temperature range or input current setting. The test function is switched off automatically after 1 minute and the instrument is working in the measuring mode Wait time (only available via interface commands, see 14, Data format UPP ) Using a pyrometer with RS485 it is possible that the connection is not fast enough to receive the pyrometer s answer to an instruction of the master. In this case a wait time can be set to slow down the data transfer (e.g.: tw = 02 at a baud rate 9600 means a wait time of 2 / 9600 sec). Settings: 00 Bit. 99 Bit 9.13 Switch contact The pyrometer is equipped with a built-in switch contact which allows to use the pyrometer as a thermo switch. This function enables the detection of a hot object in the measuring beam of the pyrometer. The contact is activated only in combination with a clear time settings auto or clear times 1 s (see 9.3 tcl / s (Clear times of the maximum value storage)). If the temperature exceeds 2 C min. or 1% of the span of the temperature sub range above the minimum temperature, the supply voltage (pin K) is connected to pin J. 16

17 IMPAC pyrometers IS 12-TSP IGA 12-TSP 10 Settings via interface and software The operating and analyzing software InfraWin is included in delivery of the pyrometer. With this software all pyrometer functions also can be used on the PC (except changing the interface or using the test current function). This section gives an overview about the functions of the software. Additionally there is a description of the individual icons in the program s help menu. Click on the F1 button after InfraWin has been loaded or click on the? in the menu bar. The following descriptions corresponds to program version 4.0. The latest version is available for free as download from the homepage Installation For installation select the setup program setup.exe from the InfraWin-CD (or the downloaded und unpacked zip file) and follow the installation instructions Program start After installation and the first program start a language must be chosen (German, English, French, Italian, Spanish. The language also can be changed in the program). On the start page the screen shows the following icons: 10.3 The start menu Opens a saved file Storage of measured values for further processing Online measurement with color bar display Online measurement with graphic display Setting of the parameters of the instrument Setting of interface, baud rate and pyrometer addresses (RS485) Time interval between two measurements Number of connected instruments (max. 2) Listing of measured or stored values in tabular form Processing of measured (stored) readings in graph form Processing of measured (stored) readings in a text file Calculation of spot sizes in various measuring distances Only if available: controls the programmable controller PI Beginning Before using the software, the serial interface connected to the pyrometer has to be selected under the Computer icon. For two devices using the RS232 interface, two PC interfaces must be used. Two devices using RS485 may be operated simultaneously by the same interface, if two different addresses have been properly entered (see 9.9, ADR (address)) Number of devices With a click on number of devices InfraWin changes to the display of 1 or 2 devices. If 2 devices are selected, always 2 windows are displayed for settings or evaluation. 17

18 IMPAC pyrometers IS 12-TSP IGA 12-TSP 10.6 Basic settings Under pyrometer parameters all preset values can be displayed and modified if necessary. If the window pyrometer parameters displayed, changing of settings on the pyrometer is blocked. The window pyrometer parameter contains all parameter settings described in 9, Parameters. Choose the correct setting for your application, the actual setting is displayed. Notes: Basic range displays the total range of the pyrometer automatically and can not be changed. If the sub range is changed the new values must be confirmed with OK. Under Material you have the possibility to store the names of different measuring objects with their emissivity values and to recall them from the list. If the pyrometer is equipped with a laser targeting light the laser button ( ) is visible. It may be turned on or off at this point. After two minutes the laser targeting light is switched off automatically. This icon opens a window that allows setting of temperature values and the hysteresis of the limit contacts. The set hysteresis is valid for both contacts. The open / save button enable to store and recall own pyrometer configurations. 1 meas. shows the current measuring temperature in the pyrometer parameters window for approx. 1 second. A click on the Test icon opens a window that allows the direct communication with the pyrometer via the interface commands (see section 14, Data format UPP ). After entering an interface command (00 is the adjusted address ex works, ms is the command reading temperature value ) and a click on Send the displayed window is opened: This window already shows the answer of the pyrometer in 1 / 10. The actual temperature reading is 1513,8 C. Len indicates the length of the answered data string, incl. Carriage Return (Chr(13)). In the lower part of the window the connection with the preset baud rate can be checked. Here the command was send 500 times with baud. It has taken 4.53 seconds and no transmission errors has occurred. 18

19 IMPAC pyrometers IS 12-TSP IGA 12-TSP 10.7 Measurement (color bar) This window displays: current temperature, graphically as color bar and numerically temperature range or adjusted sub range file size and quantity of the measured values of the current measurement emissivity the internal temperature of the instrument (T int ) minimum (T min ) and maximum values (T max ) temperature of the limit contacts The color bar display shows the span of the temperature range or the adjusted sub-range. Entering temperature values in the white fields on the left and right side of the color bar, limits for the color change of the color bar can be set. These limits can also be changed by moving the small bar with the PC mouse. The color bar displays temperatures within the two limits in green color, outside the limits in red color. Changing these values also changes the values for the limit contacts S1 and S2 (see 4.1 Pin assignment on the back of the pyrometer). The targeting light ( ) can be switched on or off, if the pyrometer is equipped with it. Emi: AutoFind: In addition, there is an input field for the emissivity in the window. If the emissivity is changed, the temperature change connected with this can be read off directly. If the true temperature of the measured object is known, you can calculate the emissivity of the measured object using the "Emi: AutoFind function: A measured temperature is displayed with the current set emissivity (in this example 100%) (here: 824 C). If you press Emi: Autofind a window will open which allows you to enter the "true" temperature. Once the temperature entry has been entered and confirmed with "OK", InfraWin will then calculate the emissivity which occurs with the new temperature. This is displayed immediately and can be used for further temperature measurement Measurement (online trend) This window displays: temperature as graphical diagram current temperature quantity of the measured values and file size of the current measurement The example shows a sample reading over the period of approx. 12 seconds with a temperature range between 800 and 1500 C. The final temperature (at the end of the reading) is C. If the pyrometer is equipped with a laser targeting light this can be switched on or off. 19

20 IMPAC pyrometers IS 12-TSP IGA 12-TSP With Mark zone a temperature range can color marked for easier recognition. Setting a temperature under Threshold prevents the recording of values above or below this temperature to keep the file size small. With Scaling trend the view of the temperature range can be limited. Note: The measuring values of measurement color bar or measurement online trend are automatically saved as "standard.i12". Should you need to edit the data later, you need to save the file as another.i12-file (start menu save as) because old values are over-written when a new measurement is taken. Files from older program versions (.i10-files) can be opened and saved as.i Output listing (analyzing) For analyzing the measured values in this field all measured data appears in a numeric list. The date beside the time gives more exactly values to see what happened on time units smaller 1 s. The value specifies the time in seconds after midnight (0:00 h). The amount of data depends on the frequency that readings were taken (settings at 10.12, Time interval between two measurements). As the amount of data increases, so does the amount of storage space required to save it. In order to save room, all.i12 data files are stored by a binary code Output.TXT file (analyzing) The same file as under Output listing may be converted into a text file and can be easily opened, for example with EXCEL. With the standard import settings EXCEL automatically formats the columns accordingly (tabulator as separators) Output trend (analyzing) The graph s curve depicts the temperature change over time within the specified temperature range. Additionally, other information appears in this window; such as recorded time (x-axis) and temperature in degrees (y-axis) as well as the time and temperature at the vertical cursor line which can be dragged with the mouse. Selecting the Trend output initially causes all the saved data to be displayed. If the data exceeds an amount that can be represented reasonably, you may Zoom in on a partial segment using the mouse (such as the segment represented in the example). Under Total you can return to the representation of the entire curve. 20 Note: The last reading is saved in the standard.i12 file and automatically appears in this form upon opening Listing or Trend output. If file open was loaded using another file, the previous file will be overwritten and replaced by the standard.i12 file.

21 IMPAC pyrometers IS 12-TSP IGA 12-TSP PC sampling rate (time interval between two measurements) This function sets a time interval. After each interval one measured value is stored on the PC. The bigger the time interval the smaller will be the stored file. This function is mainly used for long term measurements Spot size calculator The measuring distance or the spot size can be calculated in front or behind of the (adjusted) main distance of the optics. After entering the aperture, the main measuring distance and the main spot size, the input of interim values calculates spot sizes in different measuring distances. 11 Transport, packaging, storage The instrument can be damaged or destroyed if shipped incorrectly. To transport or store the instrument, please use the original box or a box padded with sufficient shock-absorbing material. For storage in humid areas or shipment overseas, the device should be placed in welded foil (ideally along with silica gel) to protect it from humidity. The pyrometer is designed for a storage temperature of -20 to 70 C with non-condensing conditions. Other kind of storage can damage or malfunction the pyrometer. 12 Maintenance 12.1 Safety Attention during pyrometer services: Should the pyrometer be integrated in a running machine process the machine has to be switched off and secured against restart before servicing the pyrometer Service The pyrometer does not have any parts which require regular service, only the lens has to be kept clean. The lens can be cleaned with a soft cloth in combination with alcohol (do not use acid solutions or dilution). Also standard cloths for cleaning glasses or photo objectives can be used Changing of optics The TS pyrometers are equipped with a focusable optics. These optics can be changed against another, e.g. if the lens is scratched or the pyrometer will be used for other measuring distances. When changing the optics a recalibration of the pyrometer is not necessary. Only the lens will be replaced for changing the focusable optics. The fixing ring has to be removed with a suitable objective wrench. After removing the old lens, put in the new one with the convex side to the front. Fix the lens with a new O-ring and the new fixing ring. For differentiation the lenses are marked with a color mark at the border of the lens and on the inside of this ring is the sticker with the optics data: Focusable optics 1: color mark red / black / red Focusable optics 2: color mark red / blue / red Focusable optics 3: color mark red / green / red Focusable optics Lens O-Ring Fixing ring Sticker with Optics data 21

Tube Analyzer LogViewer 2.3

Tube Analyzer LogViewer 2.3 Tube Analyzer LogViewer 2.3 User Manual Stand: 25.9.2015 Seite 1 von 11 Name Company Date Designed by WKS 28.02.2013 1 st Checker 2 nd Checker Version history Version Author Changes Date 1.0 Created 19.06.2015

Mehr

Newest Generation of the BS2 Corrosion/Warning and Measurement System

Newest Generation of the BS2 Corrosion/Warning and Measurement System Newest Generation of the BS2 Corrosion/Warning and Measurement System BS2 System Description: BS2 CorroDec 2G is a cable and energyless system module range for detecting corrosion, humidity and prevailing

Mehr

VGM. VGM information. HAMBURG SÜD VGM WEB PORTAL USER GUIDE June 2016

VGM. VGM information. HAMBURG SÜD VGM WEB PORTAL USER GUIDE June 2016 Overview The Hamburg Süd VGM Web portal is an application that enables you to submit VGM information directly to Hamburg Süd via our e-portal Web page. You can choose to enter VGM information directly,

Mehr

VGM. VGM information. HAMBURG SÜD VGM WEB PORTAL - USER GUIDE June 2016

VGM. VGM information. HAMBURG SÜD VGM WEB PORTAL - USER GUIDE June 2016 Overview The Hamburg Süd VGM-Portal is an application which enables to submit VGM information directly to Hamburg Süd via our e-portal web page. You can choose to insert VGM information directly, or download

Mehr

Word-CRM-Upload-Button. User manual

Word-CRM-Upload-Button. User manual Word-CRM-Upload-Button User manual Word-CRM-Upload for MS CRM 2011 Content 1. Preface... 3 2. Installation... 4 2.1. Requirements... 4 2.1.1. Clients... 4 2.2. Installation guidelines... 5 2.2.1. Client...

Mehr

Overview thermostat/ temperature controller

Overview thermostat/ temperature controller Thermostat TR-238 The Thermostat TR-238 is a electronic two-level controller for controlling of and in climate control units and vehicles. Voltage range (12V): Voltage range (24V): Control range: Hystereses:

Mehr

PROFIBUS-DP Repeater 1 to 1 and 1 to 5 with optional level converter module

PROFIBUS-DP Repeater 1 to 1 and 1 to 5 with optional level converter module LSS PROFIBUS-DP Repeater 1 to 1 and 1 to 5 with optional level converter module The LSS PROFIBUS-DP repeaters 1 to 1 and 1 to 5 are used for coupling up to six PROFIBUS bus segments in RS 485 bus technology.

Mehr

NEWSLETTER. FileDirector Version 2.5 Novelties. Filing system designer. Filing system in WinClient

NEWSLETTER. FileDirector Version 2.5 Novelties. Filing system designer. Filing system in WinClient Filing system designer FileDirector Version 2.5 Novelties FileDirector offers an easy way to design the filing system in WinClient. The filing system provides an Explorer-like structure in WinClient. The

Mehr

Electrical tests on Bosch unit injectors

Electrical tests on Bosch unit injectors Valid for Bosch unit injectors with order numbers 0 414 700 / 0 414 701 / 0 414 702 Parts Kit Magnet*: - F00H.N37.925 - F00H.N37.933 - F00H.N37.934 * For allocation to the 10-place Bosch order number,

Mehr

Electrical testing of Bosch common rail piezo injectors

Electrical testing of Bosch common rail piezo injectors Applies to generation CRI 3: Bosch 10-position order number 0 445 115 = CRI 3-16 (CRI 3.0) 1600 bar 0 445 116 = CRI 3-18 (CRI 3.2) 1800 bar 0 445 117 = CRI 3-20 (CRI 3.3) 2000 bar Tools required: Hybrid

Mehr

Application Note. Import Jinx! Scenes into the DMX-Configurator

Application Note. Import Jinx! Scenes into the DMX-Configurator Application Note Import Jinx! Scenes into the DMX-Configurator Import Jinx! Scenen into the DMX-Configurator 2 The Freeware Jinx! is an user friendly, well understandable software and furthermore equipped

Mehr

VIDEO CALL CAMERA G-VCAM-01

VIDEO CALL CAMERA G-VCAM-01 VIDEO CALL CAMERA G-VCAM-01 EN AUS GUTEM GRUND ENGLISH 07-10 2 SAFETY AND INFORMATION Safety 7 This camera is designed to transmit video and audio signals. All other uses are expressly prohibited. 7 Protect

Mehr

Snap-in switch for switches PSE, MSM and MCS 30

Snap-in switch for switches PSE, MSM and MCS 30 Product manual Snap-in switch for switches PSE, MSM and MCS 30 CONTENTS 1. PRODUCT DESCRIPTION 2. DATA AND DIMENSIONAL DRAWINGS 2.1. Technical Data 2.2. Dimensions of PSE with a Mounting Diameter 19 mm

Mehr

Datasheet. Page 1 of 7

Datasheet. Page 1 of 7 Features 20 Encoder Positions 4-way Joystick LED-Illumination high quality signal processing Benefits Tactile multi purpose application premium design Hall Effect technology Applications Multiple switch

Mehr

Electrical testing of Bosch common rail Injectors

Electrical testing of Bosch common rail Injectors Electrical testing of Bosch common rail Injectors Contents: 1. Adapter cable for Hybridtester FSA 050 (article number 0 684 010 050 / 1 687 023 571) 2. Electrical testing of Bosch common rail solenoid

Mehr

ATEX-Check list. Compiled by: Date: Signature: Acceptable practice at the determination of flash point: Closed cup according to ISO 2719

ATEX-Check list. Compiled by: Date: Signature: Acceptable practice at the determination of flash point: Closed cup according to ISO 2719 Fire and explosion hazard ATEX 137 1999/92/EG und ATEX 95 2014/34/EU Danger assessment and determination of explosion protection zone for the test space as well as the installation site ATEX-Check list

Mehr

Power supply Interference suppressed acc. to DIN EN /- 4, EN 55011, EN CI. B, power factor corrected Power factor BöSha LED driver

Power supply Interference suppressed acc. to DIN EN /- 4, EN 55011, EN CI. B, power factor corrected Power factor BöSha LED driver Operating Instructions LED Mast Double Luminaire Callisto SC DB, incl. Inclination Adjustment, Single-Chip Technology (Please, read carefully before starting operation) Version: 16.01.2017 Model 369-M

Mehr

Cameraserver mini. commissioning. Ihre Vision ist unsere Aufgabe

Cameraserver mini. commissioning. Ihre Vision ist unsere Aufgabe Cameraserver mini commissioning Page 1 Cameraserver - commissioning Contents 1. Plug IN... 3 2. Turn ON... 3 3. Network configuration... 4 4. Client-Installation... 6 4.1 Desktop Client... 6 4.2 Silverlight

Mehr

prorm Budget Planning promx GmbH Nordring Nuremberg

prorm Budget Planning promx GmbH Nordring Nuremberg prorm Budget Planning Budget Planning Business promx GmbH Nordring 100 909 Nuremberg E-Mail: support@promx.net Content WHAT IS THE prorm BUDGET PLANNING? prorm Budget Planning Overview THE ADVANTAGES OF

Mehr

Walter Buchmayr Ges.m.b.H.

Walter Buchmayr Ges.m.b.H. Seite 1/10 Chapter Description Page 1 Advantages 3 2 Performance description 4 3 Settings 5 4 Options 6 5 Technical data 7 6 Pictures 8 http://members.aon.at/buchmayrgmbh e-mail: walter.buchmayr.gmbh@aon.at

Mehr

IS 12 IGA 12 IS 12-S IGA 12-S IMPAC-Pyrometer

IS 12 IGA 12 IS 12-S IGA 12-S IMPAC-Pyrometer IS 12 IGA 12 IS 12-S IGA 12-S IMPAC-Pyrometer Operation Manual Betriebsanleitung Proven Quality IMPAC pyrometers IS 12 IS 12-S IGA 12 IGA 12-S LumaSense Technologies 2009. All rights reserved. LumaSense

Mehr

ONLINE LICENCE GENERATOR

ONLINE LICENCE GENERATOR Index Introduction... 2 Change language of the User Interface... 3 Menubar... 4 Sold Software... 5 Explanations of the choices:... 5 Call of a licence:... 7 Last query step... 9 Call multiple licenses:...

Mehr

Brand Label. Daimler Brand & Design Navigator

Brand Label. Daimler Brand & Design Navigator Daimler Brand & Design Navigator 30. März 2016 Brand Label FUSO Financial uses the FUSO letters in combination with the word Financial in the font Daimler CS for name and label. Black lettering is set

Mehr

How-To-Do. OPC-Server with MPI and ISO over TCP/IP Communication. Content. How-To-Do OPC-Server with MPI- und ISO over TCP/IP Communication

How-To-Do. OPC-Server with MPI and ISO over TCP/IP Communication. Content. How-To-Do OPC-Server with MPI- und ISO over TCP/IP Communication How-To-Do OPC-Server with MPI and ISO over TCP/IP Content OPC-Server with MPI and ISO over TCP/IP... 1 1 General... 2 1.1 Information... 2 1.2 Reference... 2 2 Procedure for the Setup of the OPC Server...

Mehr

Monitor VIS 3xx Kurzanleitung

Monitor VIS 3xx Kurzanleitung Monitor VIS 3xx Kurzanleitung 19.08.08 Art. Nr. 22261 Inhalt: 1. Spezifikationen...2 2. Tastenfunktionen...2 3. Menüführung und Einstellungen...3 Technik nach Maß Wöhler Monitor VIS 3xx 1. Spezifikationen

Mehr

Serie IN 5 plus IMPAC-Pyrometer

Serie IN 5 plus IMPAC-Pyrometer Serie IN 5 plus IMPAC-Pyrometer Operation Betriebsanleitung Manual Betriebsanleitung User Manual Proven Quality IMPAC pyrometers series IN 5 plus LumaSense Technologies 2009. All rights reserved. LumaSense

Mehr

Ingenics Project Portal

Ingenics Project Portal Version: 00; Status: E Seite: 1/6 This document is drawn to show the functions of the project portal developed by Ingenics AG. To use the portal enter the following URL in your Browser: https://projectportal.ingenics.de

Mehr

Electrical testing of Bosch common rail solenoid valve (MV) injectors

Electrical testing of Bosch common rail solenoid valve (MV) injectors Applies to MV injector, generation: -CRI 1.0 / 2.0 / 2.1 / 2.2 -CRIN 1 / 2 / 3, with K oder AK plug Bosch 10-position order number Bosch-Bestellnummer CRI: 0 445 110 xxx Bosch-Bestellnummer CRIN: 0 445

Mehr

How-To-Do. Hardware Configuration of the CC03 via SIMATIC Manager from Siemens

How-To-Do. Hardware Configuration of the CC03 via SIMATIC Manager from Siemens How-To-Do Hardware Configuration of the CC03 via SIMATIC Manager from Siemens Content Hardware Configuration of the CC03 via SIMATIC Manager from Siemens... 1 1 General... 2 1.1 Information... 2 1.2 Reference...

Mehr

ISR 12-LO IGAR 12-LO IMPAC-Pyrometer

ISR 12-LO IGAR 12-LO IMPAC-Pyrometer ISR 12-LO IGAR 12-LO IMPAC-Pyrometer Operation Manual Betriebsanleitung Proven Quality IMPAC pyrometers ISR 12-LO IGAR 12-LO LumaSense Technologies 2009. All rights reserved. LumaSense Technologies 3033

Mehr

1. General information... 2 2. Login... 2 3. Home... 3 4. Current applications... 3

1. General information... 2 2. Login... 2 3. Home... 3 4. Current applications... 3 User Manual for Marketing Authorisation and Lifecycle Management of Medicines Inhalt: User Manual for Marketing Authorisation and Lifecycle Management of Medicines... 1 1. General information... 2 2. Login...

Mehr

Attention: Give your answers to problem 1 and problem 2 directly below the questions in the exam question sheet. ,and C = [ ].

Attention: Give your answers to problem 1 and problem 2 directly below the questions in the exam question sheet. ,and C = [ ]. Page 1 LAST NAME FIRST NAME MATRIKEL-NO. Attention: Give your answers to problem 1 and problem 2 directly below the questions in the exam question sheet. Problem 1 (15 points) a) (1 point) A system description

Mehr

Can I use an older device with a new GSD file? It is always the best to use the latest GSD file since this is downward compatible to older versions.

Can I use an older device with a new GSD file? It is always the best to use the latest GSD file since this is downward compatible to older versions. EUCHNER GmbH + Co. KG Postfach 10 01 52 D-70745 Leinfelden-Echterdingen MGB PROFINET You will require the corresponding GSD file in GSDML format in order to integrate the MGB system: GSDML-Vx.x-EUCHNER-MGB_xxxxxx-YYYYMMDD.xml

Mehr

p^db=`oj===pìééçêíáåñçêã~íáçå=

p^db=`oj===pìééçêíáåñçêã~íáçå= p^db=`oj===pìééçêíáåñçêã~íáçå= Error: "Could not connect to the SQL Server Instance" or "Failed to open a connection to the database." When you attempt to launch ACT! by Sage or ACT by Sage Premium for

Mehr

Analog GSM-Gateway TRF

Analog GSM-Gateway TRF Analog GSM-Gateway TRF GSM gateway for voice- or fax transmission 1 2009 com.sat GmbH Kommunikationssysteme Schwetzinger Str. 19 D-68519 Viernheim www.comsat.de Tel: +49-(0)180-3-768837 The connecting

Mehr

Hazards and measures against hazards by implementation of safe pneumatic circuits

Hazards and measures against hazards by implementation of safe pneumatic circuits Application of EN ISO 13849-1 in electro-pneumatic control systems Hazards and measures against hazards by implementation of safe pneumatic circuits These examples of switching circuits are offered free

Mehr

Mock Exam Behavioral Finance

Mock Exam Behavioral Finance Mock Exam Behavioral Finance For the following 4 questions you have 60 minutes. You may receive up to 60 points, i.e. on average you should spend about 1 minute per point. Please note: You may use a pocket

Mehr

p^db=`oj===pìééçêíáåñçêã~íáçå=

p^db=`oj===pìééçêíáåñçêã~íáçå= p^db=`oj===pìééçêíáåñçêã~íáçå= How to Disable User Account Control (UAC) in Windows Vista You are attempting to install or uninstall ACT! when Windows does not allow you access to needed files or folders.

Mehr

rot red braun brown rot red RS-8 rot red braun brown R S V~

rot red braun brown rot red RS-8 rot red braun brown R S V~ Kleiner Ring 9 /Germany Phone: 0049 4122 / 977 381 Fax: 0049 4122 / 977 382 Sample connections: Feedback module with integrated detection of occupied tracks for the RS-feedback bus (Lenz Digital plus)

Mehr

Schalten einer kapazitiven Last mit einem Transistor

Schalten einer kapazitiven Last mit einem Transistor . Schalten einer kapazitiven Last mit einem Transistor In allgemeinen technischen Anwendungen ist das Schalten einer kapazitiven Last eher von untergeordneter Bedeutung. In Rechnersystemen (Computernetzwerken)

Mehr

Dexatek's Alexa Smart Home Skills Instruction Guide

Dexatek's Alexa Smart Home Skills Instruction Guide Dexatek's Alexa Smart Home Skills Instruction Guide Version 0.3 Author Esther Date 01/10/18 ~ 1 ~ Revision History Version Date Editor Remark 0.1 2017/04/05 Esther Initial version 0.2 2018/01/09 Esther

Mehr

IP 140-LO IMPAC-Pyrometer

IP 140-LO IMPAC-Pyrometer IP 140-LO IMPAC-Pyrometer Operation Manual Betriebsanleitung Proven Quality IMPAC pyrometer IP 140-LO LumaSense Technologies 2009. All rights reserved. LumaSense Technologies 3033 Scott Blvd. Santa Clara,

Mehr

Guidance Notes for the eservice 'Marketing Authorisation & Lifecycle Management of Medicines' Contents

Guidance Notes for the eservice 'Marketing Authorisation & Lifecycle Management of Medicines' Contents Guidance Notes for the eservice 'Marketing Authorisation & Lifecycle Management of Medicines' Contents Login... 2 No active procedure at the moment... 3 'Active' procedure... 4 New communication (procedure

Mehr

ABB i-bus EIB. EIB Power Supply Units

ABB i-bus EIB. EIB Power Supply Units ABB i-bus EIB EIB Power Supply Units Product Range Overview EIB Power Supplies ABB STOTZ-KONTAKT GmbH, 2002 - SK 029 F 02 E Product Range Overview EIB Power Supplies! EIB Power Supply, 320 ma SV/S 30.320.5!

Mehr

R&R. Ges. für Rationalisierung und Rechentechnik mbh. R&R RR-P-508 / ST106-EX1 motor-control unit

R&R. Ges. für Rationalisierung und Rechentechnik mbh. R&R RR-P-508 / ST106-EX1 motor-control unit R&R R&R RR-P-08 / ST06-EX motor-control unit housing wxhxd weight application power supply aluminum anodized approx. 6 x 8 x 8 mm housing similar to ST0 drawing DNR967 approx. 80 g motor desk control unit

Mehr

Datenblatt. Remote-I/O - u-remote UR20-4AO-UI or 4-wire connection; 16-bit resolution; 4 outputs

Datenblatt. Remote-I/O - u-remote UR20-4AO-UI or 4-wire connection; 16-bit resolution; 4 outputs 2- or 4-wire connection; 16-bit resolution; 4 outputs The analogue output module controls up to 4 analogue actuators with +/-10 V, +/-5 V, 0...10 V, 0...5 V, 2...10 V, 1...5 V, 0...20 ma or 4...20 ma with

Mehr

Non users after Cochlear Implantation in Single Sided Deafness

Non users after Cochlear Implantation in Single Sided Deafness Non users after Cochlear Implantation in Single Sided Deafness W. Pethe*, J. Langer*, S. Lissel**, K. Begall* *HNO-Klinik, AMEOS Klinikum Halberstadt **Cochlear Implant Rehabilitationszentrum Sachsen-Anhalt

Mehr

Network premium POP UP Display

Network premium POP UP Display Premium Pop Up System seamless graphic precision very compact and versatile pop-up system quick to set up at any location comes in a number of different shapes; straight, curved, wave-shaped, stair formations,

Mehr

Deceleration Technology. Rotary Dampers with high-torque range WRD-H 0607 WRD-H 0805 WRD-H 1208 WRD-H 1610 WRD-H

Deceleration Technology. Rotary Dampers with high-torque range WRD-H 0607 WRD-H 0805 WRD-H 1208 WRD-H 1610 WRD-H Rotary Dampers with high-torque range WRD-H 67 WRD-H 85 WRD-H 128 WRD-H 161 WRD-H 21 Deceleration Technology ONLINE CALCULATION AND 2D / 3D CAD DOWNLOAD M m L F Benefits Applications: - Mechanical and

Mehr

NTP Synchronisierung NTP Synchronizer

NTP Synchronisierung NTP Synchronizer Q-App: NTP Synchronisierung NTP Synchronizer Q-App zur automatischen Datums und Zeitsynchronisierung Q-App for automatic date and time synchronization Beschreibung Der Workflow hat 2 Ebenen eine Administratoren-

Mehr

Technische Information. Toolset Software und Cosworth Geräte

Technische Information. Toolset Software und Cosworth Geräte Technische Information Technical Information 07/2014 911 GT3 Cup (991) Toolset Software und Cosworth Geräte Toolset software and Cosworth devices Fahrzeug / Vehicle: 911 GT3 Cup (991) Bauteil / Part: ICD

Mehr

MECHANICAL WRENCH SERIES

MECHANICAL WRENCH SERIES MECANICAL WRENC SERIES MECANICAL WRENC SERIES The mechanical wrenches in the new Atlas Copco Saltus product line form the basis for manual tightening. They enable you to find the optimal solution for

Mehr

a) Name and draw three typical input signals used in control technique.

a) Name and draw three typical input signals used in control technique. 12 minutes Page 1 LAST NAME FIRST NAME MATRIKEL-NO. Problem 1 (2 points each) a) Name and draw three typical input signals used in control technique. b) What is a weight function? c) Define the eigen value

Mehr

Magic Figures. We note that in the example magic square the numbers 1 9 are used. All three rows (columns) have equal sum, called the magic number.

Magic Figures. We note that in the example magic square the numbers 1 9 are used. All three rows (columns) have equal sum, called the magic number. Magic Figures Introduction: This lesson builds on ideas from Magic Squares. Students are introduced to a wider collection of Magic Figures and consider constraints on the Magic Number associated with such

Mehr

Deceleration Technology. Rotary Dampers mit hohem Drehmoment WRD-H 2515 WRD-H 3015 WRD-H 4025 WRD-H

Deceleration Technology. Rotary Dampers mit hohem Drehmoment WRD-H 2515 WRD-H 3015 WRD-H 4025 WRD-H Rotary Dampers mit hohem Drehmoment WRD-H 2515 WRD-H 3015 WRD-H 4025 WRD-H 6030 Deceleration Technology ONLINE CALCULATION AND 2D / 3D CAD DOWNLOAD M m L F Benefits Material: - Aluminium and steel Applications:

Mehr

Registration of residence at Citizens Office (Bürgerbüro)

Registration of residence at Citizens Office (Bürgerbüro) Registration of residence at Citizens Office (Bürgerbüro) Opening times in the Citizens Office (Bürgerbüro): Monday to Friday 08.30 am 12.30 pm Thursday 14.00 pm 17.00 pm or by appointment via the Citizens

Mehr

Deceleration Technology. Rotary Dampers with high-torque range WRD-H 7550 WRD-H 9565 WRD-H

Deceleration Technology. Rotary Dampers with high-torque range WRD-H 7550 WRD-H 9565 WRD-H Rotary Dampers with high-torque range WRD-H 7550 WRD-H 9565 WRD-H 12070 Deceleration Technology ONLINE CALCULATION AND 2D / 3D CAD DOWNLOAD M m L F Benefits Material: - Aluminium and steel Applications:

Mehr

Quick Guide Home Network Mode

Quick Guide Home Network Mode Quick Guide Home Network Mode English > 1 German > 3 About the Home Network Mode EN Tivizen Nano & iplug normally work on their own created networks (whose SSID starts with tivizentv or iplug ) in which

Mehr

Weather forecast in Accra

Weather forecast in Accra Weather forecast in Accra Thursday Friday Saturday Sunday 30 C 31 C 29 C 28 C f = 9 5 c + 32 Temperature in Fahrenheit Temperature in Celsius 2 Converting Celsius to Fahrenheit f = 9 5 c + 32 tempc = 21

Mehr

TUNING GUIDE. ridefox.com

TUNING GUIDE. ridefox.com TUNING GUIDE ridefox.com sag setting To achieve the best performance from your FOX suspension, adjust the air pressure to attain your proper sag setting. Sag is the amount your suspension compresses under

Mehr

Digital Rockwell-Testing-Machine KB 150 R

Digital Rockwell-Testing-Machine KB 150 R Digital Rockwell-Testing-Machine KB 150 R Rockwell, ball indentation, Vickers HVT, Brinell HBT Bright LCD color display Flexible software control motorized load system with load cell, closed loop system

Mehr

Java Tools JDK. IDEs. Downloads. Eclipse. IntelliJ. NetBeans. Java SE 8 Java SE 8 Documentation

Java Tools JDK. IDEs.  Downloads. Eclipse. IntelliJ. NetBeans. Java SE 8 Java SE 8 Documentation Java Tools JDK http://www.oracle.com/technetwork/java/javase/ Downloads IDEs Java SE 8 Java SE 8 Documentation Eclipse http://www.eclipse.org IntelliJ http://www.jetbrains.com/idea/ NetBeans https://netbeans.org/

Mehr

CABLE TESTER. Manual DN-14003

CABLE TESTER. Manual DN-14003 CABLE TESTER Manual DN-14003 Note: Please read and learn safety instructions before use or maintain the equipment This cable tester can t test any electrified product. 9V reduplicated battery is used in

Mehr

How-To-Do. Communication to Siemens OPC Server via Ethernet

How-To-Do. Communication to Siemens OPC Server via Ethernet How-To-Do Communication to Siemens OPC Server via Content 1 General... 2 1.1 Information... 2 1.2 Reference... 2 2 Configuration of the PC Station... 3 2.1 Create a new Project... 3 2.2 Insert the PC Station...

Mehr

Shock pulse measurement principle

Shock pulse measurement principle Shock pulse measurement principle a [m/s²] 4.0 3.5 3.0 Roller bearing signals in 36 khz range Natural sensor frequency = 36 khz 2.5 2.0 1.5 1.0 0.5 0.0-0.5-1.0-1.5-2.0-2.5-3.0-3.5-4.0 350 360 370 380 390

Mehr

Hinweisblatt. Für den Einsatz des MAX! Heizungssteuerungssystems haben Sie zwei Alternativen: Die MAX! Einzelraumlösung und die MAX! Hauslösung.

Hinweisblatt. Für den Einsatz des MAX! Heizungssteuerungssystems haben Sie zwei Alternativen: Die MAX! Einzelraumlösung und die MAX! Hauslösung. Hinweisblatt MAX! Heizkörperthermostat BC-RT-TRX-CyG Art.-Nr. 99017 MAX! Fensterkontakt BC-SC-Rd-WM Art.-Nr. 99023 Für den Einsatz des MAX! Heizungssteuerungssystems haben Sie zwei Alternativen: Die MAX!

Mehr

Laser LightmaXX CLUB 2.0 / 4.0 SERIES

Laser LightmaXX CLUB 2.0 / 4.0 SERIES Seite 1 von 8 Laser LightmaXX CLUB 2.0 / 4.0 SERIES ACHTUNG! Seite 2 von 8 Laserschutzbestimmungen: Durch die starke Bündelung des Laserstrahls ist die gesamte Lichtenergie auf eine geringe Fläche konzentriert.

Mehr

Finite Difference Method (FDM)

Finite Difference Method (FDM) Finite Difference Method (FDM) home/lehre/vl-mhs-1-e/folien/vorlesung/2a_fdm/cover_sheet.tex page 1 of 15. p.1/15 Table of contents 1. Problem 2. Governing Equation 3. Finite Difference-Approximation 4.

Mehr

Exercise (Part V) Anastasia Mochalova, Lehrstuhl für ABWL und Wirtschaftsinformatik, Kath. Universität Eichstätt-Ingolstadt 1

Exercise (Part V) Anastasia Mochalova, Lehrstuhl für ABWL und Wirtschaftsinformatik, Kath. Universität Eichstätt-Ingolstadt 1 Exercise (Part V) Notes: The exercise is based on Microsoft Dynamics CRM Online. For all screenshots: Copyright Microsoft Corporation. The sign ## is you personal number to be used in all exercises. All

Mehr

MINI 1-Conductor/1-Pin Receptacle Terminal Blocks; 2-Conductor/2-Pin Receptacle Terminal Blocks, Series 2020

MINI 1-Conductor/1-Pin Receptacle Terminal Blocks; 2-Conductor/2-Pin Receptacle Terminal Blocks, Series 2020 X-COM @-SYSTEM MINI 1-Conductor/1-Pin Receptacle Terminal Blocks; 2-Conductor/2-Pin Receptacle Terminal Blocks, Series 2020 0.14 1 (1.5) mm 2 AWG 24 16 500 V/6 kv/3 1 13.5 A* 0.14 1 (1.5) mm 2 AWG 24 16

Mehr

Installation/setup notes

Installation/setup notes Installation/setup notes Notes applicable for the software: stute.x1d Loudspeaker Designer stute.x1a Audio Analyzer stute.scpi SCPI controller DCX.Server Install/Test our software for free (Demo) Download

Mehr

Cycling and (or?) Trams

Cycling and (or?) Trams Cycling and (or?) Trams Can we support both? Experiences from Berne, Switzerland Roland Pfeiffer, Departement for cycling traffic, City of Bern Seite 1 A few words about Bern Seite 2 A few words about

Mehr

Installation Guide WLAN Interface

Installation Guide WLAN Interface Installation Guide 7106 7206 7010 WLAN Interface EN Installation Guide This guide explains how to install the Ethernet LAN card in your label printer. You should conult your dealer or distributor for more

Mehr

VN7640 FlexRay/CAN/LIN/Ethernet Interface Quick Start Guide. Version 1.1 English/Deutsch

VN7640 FlexRay/CAN/LIN/Ethernet Interface Quick Start Guide. Version 1.1 English/Deutsch VN7640 FlexRay/CAN/LIN/Ethernet Interface Quick Start Guide Version 1.1 English/Deutsch Quick Start Guide VN7640 ENGLISH 1 ENGLISH 1.1 Installation Step by Step Procedure Please use the drivers from the

Mehr

UNIGATE CL Konfiguration mit WINGATE

UNIGATE CL Konfiguration mit WINGATE UNIGATE CL Konfiguration mit WINGATE - UNIGATE CL Configuration via WINGATE Art.-Nr.: V3928 Deutschmann Automation GmbH & Co. KG Carl-Zeiss-Str. 8 D-65520 Bad Camberg Phone: +49-(0)6434-9433-0 Hotline:

Mehr

NETWORK PREMIUM POP UP DISPLAY

NETWORK PREMIUM POP UP DISPLAY Premium Pop Up System seamless graphic precision very compact and versatile pop-up system quick to set up at any location comes in a number of different shapes; straight, curved, wave-shaped, stair formations,

Mehr

vcdm im Wandel Vorstellung des neuen User Interfaces und Austausch zur Funktionalität V

vcdm im Wandel Vorstellung des neuen User Interfaces und Austausch zur Funktionalität V vcdm im Wandel Vorstellung des neuen User Interfaces und Austausch zur Funktionalität V0.1 2018-10-02 Agenda vcdm User Interface History Current state of User Interface User Interface X-mas 2018 Missing

Mehr

Lufft UMB Sensor Overview

Lufft UMB Sensor Overview Lufft Sensor Overview Wind Radiance (solar radiation) Titan Ventus WS310 Platinum WS301/303 Gold V200A WS300 WS400 WS304 Professional WS200 WS401 WS302 Radiance (solar radiation) Radiation 2 Channel EPANDER

Mehr

19 Network Cabinet Modular, Unequipped Frame for modular upgrading or use as open rack

19 Network Cabinet Modular, Unequipped Frame for modular upgrading or use as open rack 19 Network Cabinet Modular, Unequipped Frame for modular upgrading or use as open rack page 1 of 6 Art.-No. 692515U.. 692447U unequipped basic frame for modular upgrading depth-adjustable 19 profiles with

Mehr

TravelPilot 55/65 Active Connect. Bluetooth TELEFONMENÜ /TELEPHONE MENU

TravelPilot 55/65 Active Connect. Bluetooth TELEFONMENÜ /TELEPHONE MENU TravelPilot 55/65 Active Connect Bluetooth TELEFONMENÜ /TELEPHONE MENU Inhaltsverzeichnis / Table of content 3-10 DE 11-18 EN 2 Bluetooth Telefonmenü DE 3 Start Um Ihr Mobiltelefon zusammen mit Ihrem TravelPilot

Mehr

PeakTech Zeitprogramm erstellen über Excel - Beispiel / Timed programs with Excel Example

PeakTech Zeitprogramm erstellen über Excel - Beispiel / Timed programs with Excel Example PeakTech 6180 Zeitprogramm erstellen über Excel - Beispiel / Timed programs with Excel Example 1. Excel Datei erstellen / Create Excel file Öffnen Sie Excel und erstellen eine xls-datei unter einem beliebigen

Mehr

SAMPLE EXAMINATION BOOKLET

SAMPLE EXAMINATION BOOKLET S SAMPLE EXAMINATION BOOKLET New Zealand Scholarship German Time allowed: Three hours Total marks: 24 EXAMINATION BOOKLET Question ONE TWO Mark There are three questions. You should answer Question One

Mehr

Digital Display BDD-AM10-.-SSD. For ultrasonic transducer with SSD-interface BTL.-S1... BDD-AM10-1-SSD BDD-AM10-2-SSD BDD- AM10-1 PLC/CNC BDD- AM10-2

Digital Display BDD-AM10-.-SSD. For ultrasonic transducer with SSD-interface BTL.-S1... BDD-AM10-1-SSD BDD-AM10-2-SSD BDD- AM10-1 PLC/CNC BDD- AM10-2 Digital Display BDD-AM10-.-SSD For ultrasonic transducer with SSD-interface BTL.-S1... BDD-AM10-1-SSD BDD-AM10-2-SSD direct connection BDD- AM10-1 parallel connection PLC/CNC BDD- AM10-2 Table of contents

Mehr

SETUP TOOL. Bedienungsanleitung User s Manual

SETUP TOOL. Bedienungsanleitung User s Manual Bedienungsanleitung User s Manual Bitte lesen Sie diese Anleitung vor der Inbetriebnahme sorgfältig durch Please read the instructions carefully before use 1. Eigenschaften Das Setup Tool ist ein speziell

Mehr

Bedienungsanleitung. User Manual

Bedienungsanleitung. User Manual Bedienungsanleitung Seite: -3 User Manual LightmaXX 5ive STAR LED LIG0009669-000 Page: 4-5 Lieber Kunde, vielen Dank das Sie sich für ein Produkt von LightmaXX entschieden haben. In der folgenden Anleitung

Mehr

Level 2 German, 2015

Level 2 German, 2015 91126 911260 2SUPERVISOR S Level 2 German, 2015 91126 Demonstrate understanding of a variety of written and / or visual German text(s) on familiar matters 2.00 p.m. Friday 4 December 2015 Credits: Five

Mehr

Release Notes BRICKware 7.5.4. Copyright 23. March 2010 Funkwerk Enterprise Communications GmbH Version 1.0

Release Notes BRICKware 7.5.4. Copyright 23. March 2010 Funkwerk Enterprise Communications GmbH Version 1.0 Release Notes BRICKware 7.5.4 Copyright 23. March 2010 Funkwerk Enterprise Communications GmbH Version 1.0 Purpose This document describes new features, changes, and solved problems of BRICKware 7.5.4.

Mehr

Ein Stern in dunkler Nacht Die schoensten Weihnachtsgeschichten. Click here if your download doesn"t start automatically

Ein Stern in dunkler Nacht Die schoensten Weihnachtsgeschichten. Click here if your download doesnt start automatically Ein Stern in dunkler Nacht Die schoensten Weihnachtsgeschichten Click here if your download doesn"t start automatically Ein Stern in dunkler Nacht Die schoensten Weihnachtsgeschichten Ein Stern in dunkler

Mehr

Bedienungsanleitung / Manual für il-debug_i Interface für den Debugger il_debug

Bedienungsanleitung / Manual für il-debug_i Interface für den Debugger il_debug Bedienungsanleitung / Manual für il-debug_i Interface für den Debugger il_debug Ing.Büro Stefan Lehmann Fürstenbergstraße 8a D-77756 Hausach Tel. (07831) 452 Fax (07831) 96428 E-Mail SL@iL-online.de Internet

Mehr

Technische Daten / Technisches Merkblatt (nach pren ) ECHOGRAPH 1016B (Best.-Nr )

Technische Daten / Technisches Merkblatt (nach pren ) ECHOGRAPH 1016B (Best.-Nr ) DISPLAY type of display CRT D10 size of display 72 x 60 grid 68 x 55 scale division coarse:10 fold ( and ) fine: 5 fold ( and ) acceleration voltage 10 kv MEASURING RANGES measuring length 0-100 mm; 0-500

Mehr

B/S/H/ Startfolie. B/S/H Bosch und Siemens Hausgeräte GmbH - KDT-T B/S/H Bosch und Siemens Hausgeräte GmbH KDT-T

B/S/H/ Startfolie. B/S/H Bosch und Siemens Hausgeräte GmbH - KDT-T B/S/H Bosch und Siemens Hausgeräte GmbH KDT-T Startfolie B/S/H/ 1 Product division Event Location Refrigeration Side by Side USA B/S/H Bosch und Siemens Hausgeräte GmbH KDT-T 2 Side-by-Side Appliances Models: Bosch Siemens KAN 56V** KAN58A** KA58NA**with

Mehr

Systemaufbau Blockdiagramm / System structure

Systemaufbau Blockdiagramm / System structure Systemaufbau Blockdiagramm / System structure... 1 Leuchtenanschluß / Connection of Luminaire... 2 Unterstützte DALI Kommandos / Supported DALI commands... 2 Fehlerverhalten / Failure behavior... 5 Standards

Mehr

Konfiguration von eduroam. Configuring eduroam

Konfiguration von eduroam. Configuring eduroam eduroam Windows 8.1 / 10 Konfiguration von eduroam Configuring eduroam 08.10.2018 kim.uni-hohenheim.de kim@uni-hohenheim.de Wissenschaftliche Einrichtungen und Universitäten bieten einen weltweiten Internetzugang

Mehr

Produktinformation _147PNdeen

Produktinformation _147PNdeen Produktinformation 201105_147PNdeen Neue Software für die TOUCH PC Auswuchtmaschinen BM 35 Touch, BM 45 Touch, BM 55 Touch Mit der Einführung der BM 11 Touch in die Nussbaum Produktlinie der Rad Auswuchtmaschinen

Mehr

X83-Konverter Instructions

X83-Konverter Instructions 1. General The X83-Konverter converts GAEB call-for-tender data *) directly into MS Excel format. Exchange phases D81, D83, P81, P83, X81 and X83 are supported. Processing of the tender takes place completely

Mehr

Hypex d.o.o. Alpska cesta 43, 4248 Lesce Slovenija Tel: +386 (0) Fax: +386 (0)

Hypex d.o.o. Alpska cesta 43, 4248 Lesce Slovenija Tel: +386 (0) Fax: +386 (0) MAINTENANCE INSTRUCTIONS MTJ MRJ SERIES Hypex d.o.o. Alpska cesta, Lesce Slovenija Tel: + (0) 00 Fax: + (0) 0 www.unimotion.eu email: sales@unimotion.eu www.unimotion.eu MTJ MRJ Series OVERVIEW Used symbols

Mehr

DPM_flowcharts.doc Page F-1 of 9 Rüdiger Siol :28

DPM_flowcharts.doc Page F-1 of 9 Rüdiger Siol :28 Contents F TOOLS TO SUPPORT THE DOCUMENTATION... F-2 F.1 GRAPHIC SYMBOLS AND THEIR APPLICATION (DIN 66 001)... F-2 F.1.1 Flow of control... F-3 F.1.2 Terminators and connectors... F-4 F.1.3 Lines, arrows

Mehr

Safety action Inspection of welds

Safety action Inspection of welds Knott GmbH Obinger Straße 15 D 83125 Eggstätt Receiver final custormer Ihre Nachricht Your letter Ihr Zeichen Your references Datum Date Abteilung Department Bearbeitet durch Contact person Unser Zeichen

Mehr

FACHKUNDE FüR KAUFLEUTE IM GESUNDHEITSWESEN FROM THIEME GEORG VERLAG

FACHKUNDE FüR KAUFLEUTE IM GESUNDHEITSWESEN FROM THIEME GEORG VERLAG FACHKUNDE FüR KAUFLEUTE IM GESUNDHEITSWESEN FROM THIEME GEORG VERLAG DOWNLOAD EBOOK : FACHKUNDE FüR KAUFLEUTE IM GESUNDHEITSWESEN Click link bellow and free register to download ebook: FACHKUNDE FüR KAUFLEUTE

Mehr

Lukas Hydraulik GmbH Weinstraße 39 D Erlangen. Mr. Sauerbier. Lukas Hydraulik GmbH Weinstraße 39 D Erlangen. edraulic rescue equipment

Lukas Hydraulik GmbH Weinstraße 39 D Erlangen. Mr. Sauerbier. Lukas Hydraulik GmbH Weinstraße 39 D Erlangen. edraulic rescue equipment Technical Report No. 028-7130 95685-050 of 22.02.2017 Client: Lukas Hydraulik GmbH Weinstraße 39 D-91058 Erlangen Mr. Sauerbier Manufacturing location: Lukas Hydraulik GmbH Weinstraße 39 D-91058 Erlangen

Mehr