1 Instructions for Use Adenovirus IgG ELISA Enzyme immunoassays (microtiter strips) for the qualitative and quantitative determination of IgG antibodies against Adenovirus in human serum and plasma. RE x8 2-8 C IBL GESELLSCHAFT FÜR IMMUNCHEMIE UND IMMUNBIOLOGIE MBH FLUGHAFENSTRASSE 52a D HAMBURG GERMANY Phone: +49 (0) IBL@IBL-Hamburg.com Fax: +49 (0)

2 ENGLISH 1. INTENDED USE Enzyme immunoassay for the qualitative and quantitative determination in human serum and plasma. 2. SUMMARY AND EXPLANATION The adenovirus is an ubiquitous pathogen of humans and animals. Many different serotypes can be distinguished. Adenoviruses are endemic in all populations throughout the year. The infection is spread both via airborne infection or smear infection. Adenoviruses mainly infest respiratory and intestinal mucosa, but also the cornea. They are accumulated in the epithelial cells and regional lymph nodes. Adenoviruses cause the widest variety of illnesses of the known respiratory viruses: diseases of the respiratory tract, pneumonia, swimming-pool-associated pharyngoconjunctival fever and epidemics of kerato-conjunctivitis. The intestinal form of adenoviral infection occurs mostly in children younger than one year. It is characterized by prevalence of symptoms of acute gastroenteritis and is often caused by the types 40 and 41. An acute adenoviral infection can be detected by virus isolation and/or serology. The serologic tests are particularly important because they document actual infection in the patient and can be applied to largescale epidemiologic investigations. ELISA tests measure predominantly the antibodies directed against the group-specific determinants on the hexon component. The recommended tests for measuring type-specific antibodies are hemagglutinin inhibition and serum neutralization. Because of the ubiquity of the adenoviruses and numerous cross-reactions between related serotypes, seroconversion involving a fourfold or greater rise in antibody infection is necessary to document infection. IgG is the predominant antibody class measured in the serologic tests. 3. TEST PRINCIPLE Solid phase enzyme-linked immunosorbent assay (ELISA) based on the sandwich principle. The wells are coated with antigen. Specific antibodies of the sample binding to the antigen coated wells are detected by a secondary enzyme conjugated antibody (E-Ab) specific for human IgG. After the substrate reaction the intensity of the color developed is proportional to the amount of IgG-specific antibodies detected. Results of samples can be determined directly using the standard curve. 4. WARNINGS AND PRECAUTIONS 1. For in-vitro diagnostic use only. For professional use only. 2. Before starting the assay, read the instructions completely and carefully. Use the valid version of the package insert provided with the kit. Be sure that everything is understood. For further information (clinical background, test performance, automation protocols, alternative applications, literature, etc.) please refer to the IBL-Homepage. 3. In case of severe damage of the kit package please contact IBL or your supplier in written form, latest one week after receiving the kit. Do not use damaged components in test runs, but keep safe for complaint related issues. 4. Obey lot number and expiry date. Do not mix reagents of different lots. Do not use expired reagents. 5. Follow good laboratory practice and safety guidelines. Wear lab coats, disposable latex gloves and protective glasses where necessary. 6. Reagents of this kit containing hazardous material may cause eye and skin irritations. See MATERIALS SUPPLIED and labels for details. Material Safety Data Sheets for this product are available on the IBL- Homepage or upon request directly from IBL. 7. Chemicals and prepared or used reagents have to be treated as hazardous waste according to national biohazard and safety guidelines or regulations. 8. Avoid contact with Stop solution. It may cause skin irritations and burns. 9. Some reagents contain sodium azide (NaN 3 ) as preservatives. In case of contact with eyes or skin, flush immediately with water. NaN 3 may react with lead and copper plumbing to form explosive metal azides. When disposing reagents, flush with a large volume of water to avoid azide build-up. 10. All reagents of this kit containing human serum or plasma have been tested and were found negative for HIV I/II, HBsAg and HCV. However, a presence of these or other infectious agents cannot be excluded absolutely and therefore reagents should be treated as potential biohazards in use and for disposal. 5. STORAGE AND STABILITY The kit is shipped at ambient temperature and should be stored at 2-8 C. Keep away from heat or direct sun light. The storage and stability of specimen and prepared reagents is stated in the corresponding chapters. Version 1 / / 4

3 ENGLISH The microtiter strips are stable up to the expiry date of the kit in the broken, but tightly closed bag when stored at 2 8 C. 6. SPECIMEN COLLECTION AND STORAGE Serum, Plasma (EDTA, Heparin) The usual precautions for venipuncture should be observed. It is important to preserve the chemical integrity of a blood specimen from the moment it is collected until it is assayed. Do not use grossly hemolytic, icteric or grossly lipemic specimens. Samples appearing turbid should be centrifuged before testing to remove any particulate material. Storage: 2-8 C -20 C Stability: 2 d > 2 d 7. MATERIALS SUPPLIED 4 x 2 ml CAL A-D 1 x 14 ml ENZCONJ IgG 1 x 12 x 8 MTP 1 x 14 ml TMB SUBS 1 x 14 ml TMB STOP 1 x 60 ml DILBUF 1 x 60 ml WASHBUF CONC 2 x FOIL 1 x BAG Standard A-D 1; 10; 35; 150 U/mL. Ready to use. Standard A = Negative Control Standard C = Weakly Positive Control Keep away from heat or direct sun light. Avoid repeated freeze-thaw cycles. Standard B = Cut-Off Control Standard D = Positive Control Contains IgG antibodies against Adenovirus, PBS, stabilizers. Enzyme Conjugate IgG Red colored. Ready to use. Contains anti-human IgG, conjugated to peroxidase, proteincontaining buffer, stabilizers. Microtiter Plate Break apart strips. Coated with specific antigen. TMB Substrate Solution Ready to use. Contains TMB. TMB Stop Solution Ready to use. 0.5 M H 2 SO 4. Diluent Buffer Ready to use. Contains PBS Buffer, BSA, < 0.1 % NaN 3. Wash Buffer, Concentrate (10x) Contains PBS Buffer, Tween 20. Adhesive Foil For covering of Microtiter Plate during incubation. Plastic Bag Resealable. For dry storage of non-used strips. 8. MATERIALS REQUIRED BUT NOT SUPPLIED 1. Micropipettes (Multipette Eppendorf or similar devices, < 3% CV). Volumes: 5; 50; 100; 500 µl 2. Calibrated measures 3. Tubes (1 ml) for sample dilution 4. 8-Channel Micropipettor with reagent reservoirs 5. Wash bottle, automated or semi-automated microtiter plate washing system 6. Microtiter plate reader capable of reading absorbance at 450 nm (reference wavelength nm) 7. Bidistilled or deionised water 8. Paper towels, pipette tips and timer 9. PROCEDURE NOTES 1. Any improper handling of samples or modification of the test procedure may influence the results. The indicated pipetting volumes, incubation times, temperatures and pretreatment steps have to be performed strictly according to the instructions. Use calibrated pipettes and devices only. 2. Once the test has been started, all steps should be completed without interruption. Make sure that required reagents, materials and devices are prepared ready at the appropriate time. Allow all reagents and specimens to reach room temperature (18-25 C) and gently swirl each vial of liquid reagent and sample before use. Mix reagents without foaming. 3. Avoid contamination of reagents, pipettes and wells/tubes. Use new disposable plastic pipette tips for each reagent, standard or specimen. Do not interchange caps. Always cap not used vials. Do not reuse wells/tubes or reagents. Version 1 / / 4

4 ENGLISH 4. Use a pipetting scheme to verify an appropriate plate layout. 5. Incubation time affects results. All wells should be handled in the same order and time sequences. It is recommended to use an 8-channel Micropipettor for pipetting of solutions in all wells. 6. Microplate washing is important. Improperly washed wells will give erroneous results. It is recommended to use a multichannel pipette or an automatic microplate washing system. Do not allow the wells to dry between incubations. Do not scratch coated wells during rinsing and aspiration. Rinse and fill all reagents with care. While rinsing, check that all wells are filled precisely with Wash Buffer, and that there are no residues in the wells. 7. Humidity affects the coated wells/tubes. Do not open the pouch until it reaches room temperature. Unused wells/tubes should be returned immediately to the resealed pouch including the desiccant. 10. PRE-TEST SETUP INSTRUCTIONS Preparation of Components Dilute/ Component dissolve 60 ml Wash Buffer ad 600 ml Diluent bidist. water 1:10 Remarks Storage Stability Warm up at 37 C to dissolve crystals. Mix vigorously. 2-8 C 8 w Dilution of Samples Relation Sample to be diluted with Relation Remarks Serum / Plasma generally Diluent Buffer 1:101 e.g. 5 µl µl Samples containing concentrations higher than the highest standard have to be diluted further. 11. TEST PROCEDURE Pipette 100 µl of each Standard and diluted sample into the respective wells of the Microtiter Plate. In the qualitative test only Standard B is used. Cover plate with adhesive foil. Incubate 60 min at C. Remove adhesive foil. Discard incubation solution. Wash plate 3 x with 300 µl of diluted Wash Buffer. Remove excess solution by tapping the inverted plate on a paper towel. Pipette 100 µl of Enzyme Conjugate into each well. Cover plate with new adhesive foil. Incubate 30 min at C. Remove adhesive foil. Discard incubation solution. Wash plate 3 x with 300 µl of diluted Wash Buffer. Remove excess solution by tapping the inverted plate on a paper towel. For adding of Substrate and Stop Solution use, if available, an 8-channel Micropipettor. Pipetting should be carried out in the same time intervals for Substrate and Stop Solution. Use positive displacement and avoid formation of air bubbles. Pipette 100 µl of TMB Substrate Solution into each well. Incubate 20 min at C in the dark (without adhesive foil). Stop the substrate reaction by adding 100 µl of TMB Stop Solution into each well. Briefly mix contents by gently shaking the plate. Color changes from blue to yellow. Measure optical density with a photometer at 450 nm (Reference-wavelength: nm) within 60 min after pipetting of the Stop Solution. 12. QUALITY CONTROL The test results are only valid if the test has been performed following the instructions. Moreover the user must strictly adhere to the rules of GLP (Good Laboratory Practice) or other applicable standards/laws. All standards must be found within the acceptable ranges as stated on the QC Certificate. If the criteria are not met, the run is not valid and should be repeated. Each laboratory should use known samples as further controls. In case of any deviation the following technical issues should be proven: Expiration dates of (prepared) reagents, storage conditions, pipettes, devices, incubation conditions and washing methods. It is recommended to participate at appropriate quality assessment trials. Version 1 / / 4

5 ENGLISH 13. CALCULATION OF RESULTS The obtained OD of the standards (y-axis, linear) are plotted against their concentration (x-axis, logarithmic) either on semi-logarithmic graph paper or using an automated method. A good fit is provided with cubic spline, 4 Parameter Logisitcs or Logit-Log. For the calculation of the standard curve, apply each signal of the standards (one obvious outlier of duplicates might be omitted and the more plausible single value might be used). The concentration of the samples can be read from the standard curve. The initial dilution has been taken into consideration when reading the results from the graph. Results of samples of higher predilution have to be multiplied with the dilution factor. Samples showing concentrations above the highest standard have to be diluted as described in PRE-TEST SETUP INSTRUCTIONS and reassayed. Typical Calibration Curve OD 450 nm (Example. Do not use for calculation!) Standard U/mL Mean OD A B C D INTERPRETATION OF RESULTS U/mL Interpretation The results themselves should not be the only reason for any < 8 negative therapeutical consequences. They have to be correlated to other 8-12 equivocal clinical observations and diagnostic tests. > 12 positive 15. EXPECTED VALUES In an in-house study, apparently healthy subjects showed the following results: Interpretation Ig Isotype n positive equivocal negative IgG % 1.8 % 5.4 % 16. LIMITATIONS OF THE PROCEDURE Specimen collection has a significant effect on the test results. See SPECIMEN COLLECTION AND STORAGE for details. For cross-reactivities, see PERFORMANCE. Azide and thimerosal at concentrations > 0.1 % interfere in this assay and may lead to false results. The following blood components do not have a significant Hemoglobin 8.0 mg/ml effect (+/- 15 % of expected) on the test results up to the Bilirubin 0.3 mg/ml concentrations stated below: Triglyceride 5.0 mg/ml 17. PERFORMANCE Adenovirus IgG (U/mL) Analytical Specificity (Cross Reactivity) No cross-reactivities were found to: RSV, Influenza A Precision Mean (U/mL) CV (%) Intra-Assay Inter-Assay Linearity Range (U/mL) Serial dilution up to Range (%) / Recovery % % Recovery after spiking (n = 3) Method Comparison Rel. Sensitivity > 95 % versus ELISA Rel. Specificity > 95 % Version 1 / / 4

6 DEUTSCH 1. ZWECKBESTIMMUNG Enzymimmunoassay zur qualitativen und quantitativen Bestimmung in humanem Serum und Plasma. 2. KLINISCHE BEDEUTUNG Das Adenovirus ist ein beim Menschen ubiquitär vorkommender pathogener Erreger. Bisher lassen sich 47 verschiedene Serotypen von Adenoviren beim Menschen unterscheiden. Am häufigsten werden die Typen 1-8 sowie 40 und 41 isoliert. Adenoviren sind durch ihr Vorkommen im Zellkern, durch ein gemeinsames KBR-Antigen und eine auffällige Stabilität gegenüber Umgebungseinflüssen charakterisiert. Sie sind über das ganze Jahr in allen Altersschichten endemisch. Bei der Übertragung von Adenoviren handelt es sich hauptsächlich um Schmier- und Tröpfcheninfektionen. Die Inkubationszeit beträgt 5-7 Tage. Adenoviren befallen besonders respiratorische und intestinale Schleimhäute, aber auch die Bindehaut der Augen. Sie werden in Epithelzellen und regionalen Lymphknoten angereichert. Bei Kindern im Vorschulalter ist eine Adenovirusinfektion die häufigste viral verursachte Erkrankung der Atemwege (Typen 1-5 und 7). Dabei treten vornehmlich akute Erkrankungen des oberen Respirationstraktes, aber auch Pneumonien auf. Letztere stellen die schwerste Form einer adenoviralen Infektion dar und sind besonders bei Kindern unter einem Jahr anzutreffen. Pharyngokonjunktival-Fieber und Entzündungen im Augenbereich kommen sowohl bei Kindern als auch bei Erwachsenen vor. Adenovirus 8 ist der Haupterreger der epidemischen Keratokonjunktivitis. Die nur schwer in Zellkultur vermehrbaren Typen 40 und 41 führen zur intestinalen Form einer Adenovirus-Infektion. Diese ähnelt in ihrer Symptomatik einer Gastroenteritis und tritt bevorzugt bei Kleinkindern vor dem ersten Lebensjahr auf. Eine akute Adenovirus-Infektion kann über eine Virusisolierung und/oder serologisch nachgewiesen werden. Die serologischen Tests sind von besonderem Interesse, weil sie Aufschluß über den aktuellen Infektionsverlauf im Patienten und die Epidemiologie geben. KBR und ELISA werden zur Bestimmung der Antikörper gegen die gruppenspezifischen Hexon-Komponenten durchgeführt. HHT und Serumneutralisation werden zum Nachweis typspezifischer Antikörper gegen Determinanten der Fiberantennen und des Capsids empfohlen. Aufgrund des ubiquitären Vorkommens von Adenoviren und den vielen Kreuzreaktionen zwischen den verwandten Serotypen spricht man erst bei einem vierfachen Anstieg der Antikörpertiter von einer manifestierten Adenovirus-Infektion. Mit den serologischen Tests werden hauptsächlich IgG Antikörper bestimmt. 3. TESTPRINZIP Sandwich-Enzymimmunoassay (ELISA). Die Wells der Mikrotiterstreifen sind mit Antigen beschichtet. Die spezifischen Antikörper der Probe binden an die Antigen-beschichteten Wells und werden von einem zweiten enzymmarkierten Antikörper (E-Ab), der gegen humanes IgG gerichtet ist, detektiert. Die Intensität der gebildeten Farbe nach der Substratreaktion ist proportional zur IgG-Konzentration. Die Ergebnisse der Proben können direkt anhand der Standardkurve bestimmt werden. 4. WARNHINWEISE UND VORSICHTSMASSNAHMEN 1. Nur zum In-vitro-Gebrauch. Nur für den Gebrauch durch Fachpersonal. 2. Vor der Testdurchführung sollte die Arbeitsanleitung vollständig und sorgfältig gelesen werden und verstanden worden sein. Die gültige Version aus dem Kit verwenden. Weitere Informationen (klinischer Hintergrund, Testcharakteristika, Automatisierung, alternative Anwendungen, Literatur, etc.) finden Sie auf der IBL-Homepage. 3. Im Falle einer erheblichen Beschädigung der Testpackung ist IBL bzw. der jeweilige Lieferant innerhalb einer Woche nach Empfang der Ware schriftlich zu benachrichtigen. Beschädigte Komponenten dürfen nicht zur Testdurchführung verwendet werden, sondern sollten solange aufbewahrt werden, bis der Transportschaden endgültig geregelt ist. 4. Chargen-Nummer und Verfallsdatum beachten. Es dürfen keine Reagenzien aus unterschiedlichen Chargen in einem Test verwendet werden. Verfallene Reagenzien dürfen nicht verwendet werden. 5. Gute Laborpraxis und Sicherheitsrichtlinien beachten. Je nach Bedarf sollten Laborkittel, Einmal- Latexhandschuhe und Schutzbrillen getragen werden. 6. Reagenzien dieses Kits, die Gefahrstoffe enthalten, können Reizungen der Augen und der Haut hervorrufen. Siehe Angaben in KOMPONENTEN DES KITS und auf den Etiketten. Sicherheitsdatenblätter für dieses Produkt sind auf der IBL-Homepage zum Download verfügbar oder auf Anfrage direkt von IBL erhältlich. 7. Chemikalien und vorbereitete oder gebrauchte Reagenzien sind unter Beachtung der jeweiligen nationalen Bestimmungen als Gefahrstoffabfall zu entsorgen. Version 1 / / 5

7 DEUTSCH 8. Kontakt mit Stopplösung vermeiden. Kann Hautreizungen und Verätzungen hervorrufen. 9. Einige Reagenzien enthalten Natriumazid (NaN 3 ) als Konservierungsmittel. Bei Kontakt mit den Augen oder der Haut gründlich mit Wasser abspülen. NaN 3 kann mit Blei und Kupfer explosive Azide bilden. Bei Entsorgung über das Abwasser gut spülen, um Azidanreicherung zu vermeiden. 10. Alle Reagenzien dieses Kits, die humanes Serum oder Plasma enthalten, ergaben bei der Prüfung auf HCV, HBsAg bzw. Antikörper gegen HIV I/II-Virus ein negatives Ergebnis. Trotzdem kann das Vorhandensein solcher infektiöser Erreger nicht mit absoluter Sicherheit ausgeschlossen werden. Die Reagenzien sollten deshalb wie potentiell infektiöses Material behandelt werden. 5. LAGERUNG UND HALTBARKEIT Der Kit wird bei Umgebungstemperatur angeliefert und sollte bei 2-8 C gelagert werden. Vor Hitze und direkter Sonneneinstrahlung schützen. Hinweise zur Lagerung und Haltbarkeit der Proben und vorbereiteten Reagenzien sind den entsprechenden Kapiteln zu entnehmen. Die Mikrotiterplatte ist auch nach dem Öffnen der Verpackung bis zum Verfallsdatum des Kits haltbar, wenn der Beutel sorgfältig wieder verschlossen und bei 2-8 C gelagert wird. 6. PROBENGEWINNUNG UND -AUFBEWAHRUNG Serum, Plasma (EDTA, Heparin) Die üblichen Vorsichtsmaßnahmen bei der Blutabnahme sind einzuhalten. Die chemische Integrität der Blutproben muss vom Zeitpunkt der Blutabnahme bis zur Testdurchführung erhalten bleiben. Keine hämolytischen, ikterischen oder lipämischen Proben verwenden. Getrübte Proben sollten vor der Testdurchführung zentrifugiert werden, um Partikel zu entfernen. Lagerung: 2-8 C -20 C Haltbarkeit: 2 d > 2 d 7. KOMPONENTEN DES KITS 4 x 2 ml CAL A-D 1 x 14 ml ENZCONJ IgG 1 x 12 x 8 MTP 1 x 14 ml TMB SUBS 1 x 14 ml TMB STOP 1 x 60 ml DILBUF 1 x 60 ml WASHBUF CONC 2 x FOIL 1 x BAG Standard A-D 1; 10; 35; 150 U/mL. Gebrauchsfertig. Standard A = Negativkontrolle Standard C = Schwach positive Kontrolle Vor Hitze und Sonneneinstrahlung schützen. Wiederholtes Auftauen und Einfrieren vermeiden. Standard B = Cut-Off Kontrolle Standard D = Positivkontrolle Enthält IgG Antikörper gegen Adenovirus, PBS, Stabilisatoren. Enzymkonjugat IgG Rot gefärbt. Gebrauchsfertig. Enthält anti-humanes IgG, konjugiert mit Peroxidase, Protein-Puffer, Stabilisatoren. Mikrotiterplatte Wells einzeln abbrechbar. Beschichtet mit spezifischem Antigen. TMB Substratlösung Gebrauchsfertig. Enthält TMB. TMB Stopplösung Gebrauchsfertig. 0.5 M H 2 SO 4. Verdünnungspuffer Gebrauchsfertig. Enthält PBS Puffer, BSA, < 0.1 % NaN 3. Waschpuffer, Konzentrat (10x) Enthält PBS Puffer, Tween 20. Haftklebefolie Zur Abdeckung der Mikrotiterplatte während der Inkubation. Plastikbeutel Wiederverschließbar. Zur Aufbewahrung der nicht gebrauchten Streifen. 8. ZUSÄTZLICHES MATERIAL (NICHT IM KIT ENTHALTEN) 1. Mikropipetten (Multipette Eppendorf oder ähnliche, < 3% VK). Volumina: 5; 50; 100; 500 µl 2. Messzylinder 3. Röhrchen (1 ml) zur Probenverdünnung 4. 8-Kanal Mikropipette mit Reagenziengefäßen 5. Waschflasche, automatisches oder halbautomatisches Waschsystem für Mikrotiterplatten 6. Messgerät für Mikrotiterplatten zur Messung der Absorption bei 450 nm (Referenzwellenlänge nm) Version 1 / / 5

8 DEUTSCH 7. Bidest. Wasser 8. Papiertücher, Pipettenspitzen, Stoppuhr 9. HINWEISE ZUR TESTDURCHFÜHRUNG 1. Fehler bei der Handhabung der Proben oder Abweichungen von der beschriebenen Testdurchführung können die Ergebnisse verfälschen. Die angegebenen Pipettiervolumina, Inkubationszeiten, Temperaturen und Vorbereitungsschritte sind unbedingt gemäß Arbeitsanleitung einzuhalten. Nur kalibrierte Pipetten und Geräte verwenden. 2. Sobald mit der Testdurchführung begonnen wird, sollten alle Arbeitsschritte ohne Unterbrechung durchgeführt werden. Es ist sicherzustellen, dass alle benötigten Reagenzien, Geräte und Hilfsmittel zur rechten Zeit zur Verfügung stehen. Alle Reagenzien und Proben müssen auf Raumtemperatur (18-25 C) gebracht und vor Gebrauch vorsichtig ohne Schaumbildung gemischt werden. 3. Kontaminationen der Reagenzien, Pipetten und Wells/Röhrchen sind zu vermeiden. Neue Einmal- Pipettenspitzen für jedes Reagenz, jeden Standard und jede Probe verwenden. Die Deckel der Fläschchen nicht vertauschen. Nicht benötigte Fläschchen immer verschlossen halten. Wells/Röhrchen oder Reagenzien dürfen nicht wiederverwendet werden. 4. Es sollte ein Pipettierschema verwendet werden um die Identifikation der Standards und Proben auf der Platte sicherzustellen. 5. Die Inkubationszeiten beeinflussen die Ergebnisse. Bei jedem Pipettierschritt sollten alle Wells in der gleichen Reihenfolge und im gleichen Zeittakt behandelt werden. Die Verwendung einer 8-Kanal- Mikropipette zum Pipettieren in alle Wells wird empfohlen. 6. Die korrekte Durchführung der Waschschritte ist entscheidend. Ungenügend gewaschene Wells ergeben falsche Ergebnisse. Die Verwendung einer Multikanalpipette oder eines automatischen Waschsystems für Mikrotiterplatten wird empfohlen. Zwischen den Inkubationen die Wells nicht austrocknen lassen. Beim Waschen und Ausschütteln dürfen die beschichteten Wells nicht beschädigt werden. Alle Reagenzien müssen daher mit Vorsicht pipettiert werden. Beim Waschvorgang ist es wichtig, dass alle Wells vollständig und gleichmäßig mit Waschpuffer gefüllt werden und nach dem Ausschütteln kein Rückstand an Flüssigkeit zurückbleibt. 7. Feuchtigkeit beeinflusst die beschichteten Wells/Röhrchen. Verpackung nicht öffnen bevor Raumtemperatur erreicht ist. Nicht benötigte Wells/Röhrchen sofort in den wiederverschließbaren Beutel mit Trockenmittel zurückgeben. 10. TESTVORBEREITUNGEN Vorbereitung der Komponenten Verd./ rekonst. Komponente 60 ml Waschpuffer ad 600 ml Diluent bidest. Wasser 1:10 Bemerkungen Auf 37 C erwärmen, um Kristalle aufzulösen. Gründlich mischen. 2-8 C Verhältnis Lagerung Haltbarkeit 8 w Probenverdünnung zu Verhältnis Probe mit Bemerkungen verdünnen Serum / Plasma immer Verdünnungspuffer 1:101 z.b. 5 µl µl Proben mit Konzentrationen über dem höchsten Standard müssen weiter verdünnt werden. 11. TESTDURCHFÜHRUNG Je 100 µl von jedem Standard und jeder verdünnten Probe in die entsprechenden Wells der Mikrotiterplatte pipettieren. Im qualitativen Assay wird nur Standard B verwendet. Platte mit Haftklebefolie abdecken. 60 min bei C inkubieren. Folie entfernen. Inkubationslösung verwerfen. Platte 3 x mit je 300 µl verdünntem Waschpuffer waschen. Restliche Flüssigkeit auf Papiertüchern ausklopfen. 100 µl Enzymkonjugat in jedes Well pipettieren. Platte mit neuer Folie abdecken. 30 min bei C inkubieren. Version 1 / / 5

9 DEUTSCH Folie entfernen. Inkubationslösung verwerfen. Platte 3 x mit je 300 µl verdünntem Waschpuffer waschen. Restliche Flüssigkeit auf Papiertüchern ausklopfen. Substrat- und Stopplösung möglichst mit einer 8-Kanal-Pipette pipettieren und Substrat- und Stopplösung in denselben Zeitintervallen zugeben. Mit positivem Vorhub pipettieren, um die Bildung von Luftbläschen zu vermeiden. 100 µl TMB Substratlösung in jedes Well pipettieren. 20 min bei RT (18-25 C) im Dunkeln inkubieren. (Ohne Haftklebefolie.) Die Substratreaktion durch Zugabe von 100 µl TMB Stopplösung in jedes Well stoppen. Platte kurz schütteln. Die Farbe wechselt von blau nach gelb. Die optische Dichte mit einem Photometer innerhalb von 60 min nach Zugabe der Stopplösung bei 450 nm messen (Referenzwellenlänge: nm). 12. QUALITÄTSKONTROLLE Die Testergebnisse sind nur gültig, wenn der Rest gemäß der vorliegenden Arbeitsanleitung abgearbeitet wurde. Ferner muss der Anwender die GLP- Regeln (Good Laboratory Practice) und andere einschlägige Normen und Gesetze beachten. Alle Standards des Kits müssen innerhalb der Akzeptanzbereiche, die auf dem QC-Zertifikat angegeben sind, gefunden werden. Wenn die Kriterien nicht erfüllt sind, sind die Ergebnisse ungültig und der Test sollte wiederholt werden. Jedes Labor sollte darüber hinaus eigene bekannte Proben als weitere Kontrollen mitführen. Es wird empfohlen, an den einschlägigen Ringversuchen teilzunehmen. Bei Abweichungen sind die folgenden Fehlermöglichkeiten zu überprüfen: Haltbarkeit der (vorbereiteten) Reagenzien, Lagerungsbedingungen, Pipetten, Geräte und Hilfsmittel, Inkubationsbedingungen und Waschmethoden. 13. TESTAUSWERTUNG Die erhaltenen OD der Standards (y-achse, linear) gegen deren Konzentration (x-achse, logarithmisch) auftragen, entweder auf semi-logarithmischem Papier oder durch ein entsprechendes Computerprogramm. Bei Verwendung eines Computerprogramms werden die Cubic-Spline-Methode, 4-Parameter-Analyse (linlog) oder Logit-Log-Berechnung empfohlen. Zur Berechnung der Standardkurve sollten alle Werte der Standards verwendet werden (bei Doppelwerten kann ein offensichtlicher Ausreißerwert eliminiert und stattdessen der plausiblere Einzelwert verwendet werden). Die Konzentrationen der Proben können von der Standardkurve abgelesen werden. Die standardmäßig durchzuführende Verdünnung ist in dem oben beschriebenen Auswerteverfahren bereits berücksichtigt. Wenn Proben anders oder weiter verdünnt wurden, müssen die Ergebnisse mit dem entsprechenden Verdünnungsfaktor multipliziert werden. Proben, die oberhalb des höchsten Standards gemessen werden, müssen wie in TESTVORBEREITUNGEN beschrieben verdünnt und erneut analysiert werden. Typische Standardkurve OD 450 nm (Beispiel. Nicht zur Testauswertung verwenden!) Standard U/mL Mittelwert OD A B C D Adenovirus IgG (U/mL) 14. INTERPRETATION DER ERGEBNISSE U/mL Interpretation Therapeutische Konsequenzen sollten nicht allein aufgrund der < 8 negativ mit diesem Test ermittelten Werte getroffen werden, sondern nur 8-12 grenzwertig unter Berücksichtigung aller klinischen Beobachtungen und > 12 positiv weiterer diagnostischer Mittel. Version 1 / / 5

10 DEUTSCH 15. NORMWERTE In einer eigenen Studie zeigten offensichtlich gesunde Probanden die folgenden Ergebnisse: Interpretation Ig Isotyp n positiv grenzwertig negativ IgG % 1.8 % 5.4 % 16. GRENZEN DES VERFAHRENS Die korrekte Durchführung der Probengewinnung ist entscheidend für die Testergebnisse. Näheres siehe PROBENGEWINNUNG UND -LAGERUNG. Angaben zu Kreuzreaktivitäten sind im Kapitel TESTCHARAKTERISTIKA zu finden. Azid und Thimerosal in Konzentrationen > 0.1 % stören im Assay und führen evtl. zu falschen Ergebnissen. Die folgenden Blutbestandteile haben bis zu der Hämoglobin 8.0 mg/ml angegebenen Konzentration keinen signifikanten Einfluss Bilirubin 0.3 mg/ml auf die Testergebnisse (+/- 15 %): Triglyceride 5.0 mg/ml 17. TESTCHARAKTERISTIKA Analytische Spezifität Es wurden keine Kreuzreaktionen (Kreuzreaktivität) gefunden gegen: RSV, Influenza A Präzision Mittelwert (U/mL) VK (%) Intra-Assay Inter-Assay Bereich (U/mL) Höchste Bereich (%) Linearität Verdünnungsstufe / Wiederfindung % % Wiederfindung nach Spiken (n = 3) Methodenvergleich Rel. Sensitivität > 95 % versus ELISA Rel. Spezifität > 95 % Version 1 / / 5

11 Symbols / Symbole / Symbôles / Símbolos / Símbolos / Σύµβολα REF LOT CONC LYO IVD Cat.-No.: / Kat.-Nr.: / No.- Cat.: / Cat.-No.: / N.º Cat.: / N. Cat.: / Αριθµός-Κατ.: Lot-No.: / Chargen-Bez.: / No. Lot: / Lot-No.: / Lote N.º: / Lotto n.: / Αριθµός -Παραγωγή: Use by: / Verwendbar bis: / Utiliser à: / Usado por: / Usar até: / Da utilizzare entro: / Χρησιµοποιείται από: No. of Tests: / Kitgröße: / Nb. de Tests: / No. de Determ.: / N.º de Testes: / Quantità dei tests: / Αριθµός εξετάσεων: Concentrate / Konzentrat / Concentré / Concentrar / Concentrado / Concentrato / Συµπύκνωµα Lyophilized / Lyophilisat / Lyophilisé / Liofilizado / Liofilizado / Liofilizzato / Λυοφιλιασµένο In Vitro Diagnostic Medical Device. / In-vitro-Diagnostikum. / Appareil Médical pour Diagnostics In Vitro. / Dispositivo Médico para Diagnóstico In Vitro. / Equipamento Médico de Diagnóstico In Vitro. / Dispositivo Medico Diagnostico In vitro. / Ιατρική συσκευή για In-Vitro ιάγνωση. Evaluation kit. / Nur für Leistungsbewertungszwecke. / Kit pour évaluation. / Juego de Reactivos para Evaluació. / Kit de avaliação. / Kit di evaluazione. / Κιτ Αξιολόγησης. Read instructions before use. / Arbeitsanleitung lesen. / Lire la fiche technique avant emploi. / Lea las instrucciones antes de usar. / Ler as instruções antes de usar. / Leggere le istruzioni prima dell uso. / ιαβάστε τις οδηγίες πριν την χρήση. Keep away from heat or direct sun light. / Vor Hitze und direkter Sonneneinstrahlung schützen. / Garder à l abri de la chaleur et de toute exposition lumineuse. / Manténgase alejado del calor o la luz solar directa. / Manter longe do calor ou luz solar directa. / Non esporre ai raggi solari. / Να φυλάσσεται µακριά από θερµότητα και άµεση επαφή µε το φως του ηλίου. Store at: / Lagern bei: / Stocker à: / Almacene a: / Armazenar a: / Conservare a: / Αποθήκευση στους: Manufacturer: / Hersteller: / Fabricant: / Productor: / Fabricante: / Fabbricante: / Παραγωγός: Caution! / Vorsicht! / Attention! / Precaución! / Cuidado! / Attenzione! / Προσοχή! Symbols of the kit components see MATERIALS SUPPLIED. Die Symbole der Komponenten sind im Kapitel KOMPONENTEN DES KITS beschrieben. Voir MATERIEL FOURNI pour les symbôles des composants du kit. Símbolos de los componentes del juego de reactivos, vea MATERIALES SUMINISTRADOS. Para símbolos dos componentes do kit ver MATERIAIS FORNECIDOS. Per i simboli dei componenti del kit si veda COMPONENTI DEL KIT. Για τα σύµβολα των συστατικών του κιτ συµβουλευτείτε το ΠΑΡΕΧΟΜΕΝΑ ΥΛΙΚΑ. IBL AFFILIATES WORLDWIDE IBL-Hamburg GmbH Flughafenstr. 52A, D Hamburg, Germany Tel.: + 49 (0) Fax: ibl@ibl-hamburg.com WEB: LIABILITY: Complaints will only be accepted in written and if all details of the test performance and results are included (complaint form available from IBL or supplier). Any modification of the test procedure or exchange or mixing of components of different lots could negatively affect the results. These cases invalidate any claim for replacement. Regardless, in the event of any claim, the manufacturer s liability is not to exceed the value of the test kit. Any damage caused to the kit during transportation is not subject to the liability of the manufacturer. Symbols Version 2.01 /

Transglutaminase IgA ELISA

Transglutaminase IgA ELISA Instructions for Use Transglutaminase IgA ELISA Enzyme immunoassays (microtiter strips) for the qualitative and quantitative determination of IgA antibodies against Transglutaminase in human serum and

Mehr

Echinococcus IgG ELISA

Echinococcus IgG ELISA Instructions for Use Echinococcus IgG ELISA Enzyme immunoassay for the qualitative and quantitative determination of IgG antibodies against Echinococcus in human serum and plasma. RE56201 12x8 2-8 C I

Mehr

Scl-70-Ab ELISA. Enzymimmunoassay zur qualitativen und quantitativen Bestimmung von IgG-Antikörpern gegen Scl-70 in humanem Serum.

Scl-70-Ab ELISA. Enzymimmunoassay zur qualitativen und quantitativen Bestimmung von IgG-Antikörpern gegen Scl-70 in humanem Serum. Arbeitsanleitung Scl-70-Ab ELISA Enzymimmunoassay zur qualitativen und quantitativen Bestimmung von IgG-Antikörpern gegen Scl-70 in humanem Serum. RE75251 12x8 2-8 C I B L I N T E R N A T I O N A L G M

Mehr

Testosterone ELISA. Enzymimmunoassay zur quantitativen Bestimmung von Testosteron in humanem Serum und Plasma.

Testosterone ELISA. Enzymimmunoassay zur quantitativen Bestimmung von Testosteron in humanem Serum und Plasma. Arbeitsanleitung Testosterone ELISA Enzymimmunoassay zur quantitativen Bestimmung von Testosteron in humanem Serum und Plasma. RE52151 96 2-8 C I B L I N T E R N A T I O N A L G M B H Flughafenstrasse

Mehr

Cardiolipin IgM ELISA

Cardiolipin IgM ELISA Arbeitsanleitung Cardiolipin IgM ELISA Enzymimmunoassay zur qualitativen und quantitativen Bestimmung von IgM Antikörpern gegen Cardiolipin in humanem Serum. RE75021 12x8 2-8 C I B L I N T E R N A T I

Mehr

Gliadin IgA ELISA. Enzymimmunoassay zur qualitativen und quantitativen Bestimmung von IgA-Antikörpern gegen Gliadin in humanem Serum.

Gliadin IgA ELISA. Enzymimmunoassay zur qualitativen und quantitativen Bestimmung von IgA-Antikörpern gegen Gliadin in humanem Serum. Arbeitsanleitung Gliadin IgA ELISA Enzymimmunoassay zur qualitativen und quantitativen Bestimmung von IgA-Antikörpern gegen Gliadin in humanem Serum. RE75701 12x8 2-8 C I B L I N T E R N A T I O N A L

Mehr

Brucella IgA ELISA. 12x8

Brucella IgA ELISA. 12x8 Arbeitsanleitung Brucella IgA ELISA Enzymimmunoassays (Mikrotiterstreifen) zur qualitativen und quantitativen Bestimmung von IgA Antikörpern gegen Brucella in humanem Serum und Plasma. RE56831 12x8 2-8

Mehr

Influenza B IgM ELISA

Influenza B IgM ELISA Instructions for Use Influenza B IgM ELISA Enzyme immunoassays (microtiter strips) for the qualitative and quantitative determination of IgM antibodies against Influenza B virus in human serum and plasma.

Mehr

Mumps Virus IgG ELISA

Mumps Virus IgG ELISA Instructions for Use Mumps Virus IgG ELISA Enzyme immunoassays (microtiter strips) for the qualitative and quantitative determination of IgG antibodies against Mumps Virus in human serum and plasma. RE56641

Mehr

Spermatozoa-Ab ELISA

Spermatozoa-Ab ELISA Arbeitsanleitung Spermatozoa-Ab ELISA Enzymimmunoassay für die quantitative Bestimmung von Autoantikörpern gegen Oberflächenantigene von Spermatozoen in humanem Serum. RE52029 96 I B L I N T E R N A T

Mehr

I B L I N T E R N A T I O N A L G M B H

I B L I N T E R N A T I O N A L G M B H Instructions for Use Tetanus IgG ELISA Enzyme immunoassays (microtiter strips) for the quantitative determination of IgG antibodies against Tetanus in human serum and plasma. RE56901 12x8 2-8 C I B L I

Mehr

CEA ELISA. Enzyme immunoassay for the in-vitro-diagnostic quantitative determination of carcino-embryonic antigen in human serum.

CEA ELISA. Enzyme immunoassay for the in-vitro-diagnostic quantitative determination of carcino-embryonic antigen in human serum. Instructions for Use CEA ELISA Enzyme immunoassay for the in-vitro-diagnostic quantitative determination of carcino-embryonic antigen in human serum. RE59101 96 2-8 C I B L I N T E R N A T I O N A L G

Mehr

Influenza A IgM ELISA

Influenza A IgM ELISA Instructions for Use Influenza A IgM ELISA Enzyme immunoassays (microtiter strips) for the qualitative and quantitative determination of IgM antibodies against Influenza A virus in human serum and plasma.

Mehr

Candida albicans IgG ELISA

Candida albicans IgG ELISA Instructions for Use Candida albicans IgG ELISA Enzyme immunoassays for the qualitative and quantitative determination of IgG antibodies against Candida albicans in human serum and plasma. RE56171 12x8

Mehr

I B L I N T E R N A T I O N A L G M B H

I B L I N T E R N A T I O N A L G M B H Arbeitsanleitung Tetanus IgG ELISA Enzymimmunoassays (Mikrotiterstreifen) für die quantitative Bestimmung von IgG Antikörpern gegen Tetanus in humanem Serum und Plasma. RE56901 12x8 2-8 C I B L I N T E

Mehr

SS-A/Ro-Ab ELISA. Enzymimmunoassay zur qualitativen und quantitativen Bestimmung von IgG-Antikörpern gegen SS-A/Ro in humanem Serum.

SS-A/Ro-Ab ELISA. Enzymimmunoassay zur qualitativen und quantitativen Bestimmung von IgG-Antikörpern gegen SS-A/Ro in humanem Serum. Arbeitsanleitung SS-A/Ro-Ab ELISA Enzymimmunoassay zur qualitativen und quantitativen Bestimmung von IgG-Antikörpern gegen SS-A/Ro in humanem Serum. RE75231 12x8 2-8 C I B L I N T E R N A T I O N A L G

Mehr

Helicobacter pylori IgA ELISA

Helicobacter pylori IgA ELISA Arbeitsanleitung Helicobacter pylori IgA ELISA Enzymimmunoassay für die qualitative und quantitative Bestimmung von IgA Antikörpern gegen Helicobacter pylori in humanem Serum und Plasma. RE56371 12x8 2-8

Mehr

Bordetella pertussis IgA ELISA

Bordetella pertussis IgA ELISA Arbeitsanleitung Bordetella pertussis IgA ELISA Enzyme immunoassays for the qualitative and quantitative determination of IgA antibodies against Bordetella pertussis in human serum and plasma. RE56131

Mehr

Bordetella pertussis IgM ELISA

Bordetella pertussis IgM ELISA Instructions for Use Bordetella pertussis IgM ELISA Enzyme immunoassays for the qualitative and quantitative determination of IgM antibodies against Bordetella pertussis in human serum and plasma. RE56151

Mehr

Thyroid Peroxidase-Ab ELISA

Thyroid Peroxidase-Ab ELISA Arbeitsanleitung Thyroid Peroxidase-Ab ELISA Enzymimmunoassay zur qualitativen und quantitativen Bestimmung von IgG-Antikörpern gegen Schilddrüsen-Peroxidase (Thyreoidale Peroxidase, TPO) in humanem Serum.

Mehr

Epstein Barr Virus (EBNA-1) IgM ELISA

Epstein Barr Virus (EBNA-1) IgM ELISA Instructions for Use Epstein Barr Virus (EBNA-1) IgM ELISA Enzyme immunoassays (microtiter strips) for the qualitative and quantitative determination of IgM antibodies against the of Epstein Barr Virus

Mehr

Tetanus IgG ELISA. 12x8

Tetanus IgG ELISA. 12x8 Arbeitsanleitung Tetanus IgG ELISA Enzymimmunoassays (Mikrotiterstreifen) für die quantitative Bestimmung von IgG Antikörpern gegen Tetanus in humanem Serum und Plasma. RE56901 12x8 2-8 C I B L I N T E

Mehr

Diphtheria IgG ELISA

Diphtheria IgG ELISA Instructions for Use Diphtheria IgG ELISA Enzyme immunoassay for the quantitative determination of IgG antibodies against Diphtheria in human serum and plasma. RE56191 12x8 2-8 C I B L I N T E R N A T

Mehr

Gliadin IgM ELISA. IBL GESELLSCHAFT FÜR IMMUNCHEMIE UND IMMUNBIOLOGIE MBH FLUGHAFENSTRASSE 52a D HAMBURG GERMANY

Gliadin IgM ELISA. IBL GESELLSCHAFT FÜR IMMUNCHEMIE UND IMMUNBIOLOGIE MBH FLUGHAFENSTRASSE 52a D HAMBURG GERMANY Instructions for Use Gliadin IgM ELISA Enzyme immunoassays (microtiter strips) for the qualitative and quantitative determination of IgM antibodies against Gliadin in human serum and plasma. RE56331 12x8

Mehr

Testosterone Saliva ELISA

Testosterone Saliva ELISA Arbeitsanleitung Testosterone Saliva ELISA Enzymimmunoassay zur quantitativen Bestimmung von freiem Testosteron in humanem Saliva. RE52631 96 2-8 C I B L I N T E R N A T I O N A L G M B H Flughafenstrasse

Mehr

Aspergillus fumigatus IgM ELISA

Aspergillus fumigatus IgM ELISA Arbeitsanleitung Aspergillus fumigatus IgM ELISA Enzymimmunoassays zur qualitativen und quantitativen Bestimmung von IgM Antikörpern gegen Aspergillus fumigatus in humanem Serum und Plasma. RE56121 12x8

Mehr

Cytomegalovirus IgG ELISA

Cytomegalovirus IgG ELISA Arbeitsanleitung Cytomegalovirus IgG ELISA Enzymimmunoassay für die quantitative Bestimmung von IgG Antikörpern gegen das Cytomegalovirus in humanem Serum und Plasma. RE58311 96 2-8 C I B L I N T E R N

Mehr

DHEA-S ELISA. Enzymimmunoassay zur quantitativen Bestimmung von DHEA-S in humanem Serum oder Plasma. 1 x 96

DHEA-S ELISA. Enzymimmunoassay zur quantitativen Bestimmung von DHEA-S in humanem Serum oder Plasma. 1 x 96 Arbeitsanleitung DHEA-S ELISA Enzymimmunoassay zur quantitativen Bestimmung von DHEA-S in humanem Serum oder Plasma. RE52181 1 x 96 2-8 C I B L I N T E R N A T I O N A L G M B H Flughafenstrasse 52a Phone:

Mehr

Cortisol ELISA. Enzymimmunoassay zur quantitativen Bestimmung von Cortisol in humanem Serum und Plasma.

Cortisol ELISA. Enzymimmunoassay zur quantitativen Bestimmung von Cortisol in humanem Serum und Plasma. Arbeitsanleitung Cortisol ELISA Enzymimmunoassay zur quantitativen Bestimmung von Cortisol in humanem Serum und Plasma. RE52061 96 2-8 C I B L I N T E R N A T I O N A L G M B H Flughafenstrasse 52a Phone:

Mehr

Mycoplasma pneumoniae IgM ELISA

Mycoplasma pneumoniae IgM ELISA Instructions for Use Mycoplasma pneumoniae IgM ELISA Enzyme immunoassays (microtiter strips) for the quantitative determination of IgM antibodies against Mycoplasma pneumoniae in human serum and plasma.

Mehr

Free Testosterone ELISA

Free Testosterone ELISA Arbeitsanleitung Free Testosterone ELISA Enzymimmunoassay zur quantitativen In-vitro-Bestimmung von freiem Testosteron in humanem Serum. DB52181 96 2-8 C I B L I N T E R N A T I O N A L G M B H Flughafenstrasse

Mehr

Influenza A IgA ELISA

Influenza A IgA ELISA Arbeitsanleitung Influenza A IgA ELISA Enzymimmunoassays (Mikrotiterstreifen) zur qualitativen und quantitativen Bestimmung von IgA Antikörpern gegen Influenza A Virus in humanem Serum und Plasma. RE56501

Mehr

Parainfluenza 1/2/3 IgG ELISA

Parainfluenza 1/2/3 IgG ELISA Instructions for Use Parainfluenza 1/2/3 IgG ELISA Enzyme immunoassays (microtiter strips) for the qualitative and quantitative determination of IgG antibodies against Parainfluenza virus 1,2 and 3 in

Mehr

Treponema pallidum Screen ELISA

Treponema pallidum Screen ELISA Arbeitsanleitung Treponema pallidum Screen ELISA Enzymimmunoassay zur qualitativen Bestimmung von Treponema pallidum spezifischen Gesamtantikörper (TAB) in humanem Serum oder Plasma. RE58821 RE58829 12x8

Mehr

Modified Gliadin Peptide (MGP) IgG ELISA

Modified Gliadin Peptide (MGP) IgG ELISA Arbeitsanleitung Modified Gliadin Peptide (MGP) IgG ELISA Enzymimmunoassay zur qualitativen und quantitativen Bestimmung von IgG-Antikörpern gegen MPG in humanem Serum. RE75751 96 2-8 C I B L I N T E R

Mehr

Borrelia 14kD + OspC IgM ELISA

Borrelia 14kD + OspC IgM ELISA Instructions for Use Borrelia 14kD + OspC IgM ELISA Enzyme immunoassay for the semi-quantitative determination or qualitative detection in human serum and CSF. RE57211 96 2-8 C I B L I N T E R N A T I

Mehr

Chromogranin A ELISA

Chromogranin A ELISA Arbeitsanleitung Chromogranin A ELISA Enzymimmunoassay zur quantitativen Bestimmung von human Chromogranin A in humanem Serum und Plasma (EDTA). RE53971 96 2-8 C I B L I N T E R N A T I O N A L G M B H

Mehr

Free Testosterone ELISA

Free Testosterone ELISA Arbeitsanleitung Free Testosterone ELISA Enzymimmunoassay zur quantitativen Bestimmung von freiem Testosteron in humanem Serum. DB52181 96 2-8 C I B L I N T E R N A T I O N A L G M B H Flughafenstrasse

Mehr

Herpes simplex virus 1/2 IgM ELISA

Herpes simplex virus 1/2 IgM ELISA Arbeitsanleitung Herpes simplex virus 1/2 IgM ELISA Enzymimmunoassays (Mikrotiterstreifen) zur qualitativen und quantitativen Bestimmung von IgM Antikörpern gegen das Herpes simplex Virus 1 und 2 in humanem

Mehr

DHEA ELISA. Enzymimmmunoassay für die quantitative in-vitro-diagnostische Bestimmung von DHEA in humanem Serum.

DHEA ELISA. Enzymimmmunoassay für die quantitative in-vitro-diagnostische Bestimmung von DHEA in humanem Serum. Arbeitsanleitung DHEA ELISA Enzymimmmunoassay für die quantitative in-vitro-diagnostische Bestimmung von DHEA in humanem Serum. RE52221 96 2-8 C I B L I N T E R N A T I O N A L G M B H Flughafenstrasse

Mehr

Cytomegalovirus IgM ELISA

Cytomegalovirus IgM ELISA Arbeitsanleitung Cytomegalovirus IgM ELISA Enzymimmunoassay für die quantitative Bestimmung von IgM Antikörpern gegen das Cytomegalovirus in humanem Serum und Plasma. RE58321 96 2-8 C I B L I N T E R N

Mehr

Yersinia enterocolitica IgG ELISA

Yersinia enterocolitica IgG ELISA Arbeitsanleitung Yersinia enterocolitica IgG ELISA Enzymimmunoassays (Mikrotiterstreifen) zur qualitativen und quantitativen Bestimmung von IgG Antikörpern gegen Yersinia enterocolitica in humanem Serum

Mehr

Rubella virus IgG ELISA

Rubella virus IgG ELISA Arbeitsanleitung Rubella virus IgG ELISA Enzymimmunoassay für die quantitative Bestimmung von IgG Antikörpern gegen das Röteln Virus in humanem Serum und Plasma. RE58341 96 2-8 C I B L I N T E R N A T

Mehr

Progesterone Luminescence Immunoassay

Progesterone Luminescence Immunoassay Instructions for Use Progesterone Luminescence Immunoassay Luminescence immunoassay for the in-vitro-diagnostic quantitative determination of progesterone in human saliva. RE62021 96 2-8 C I B L I N T

Mehr

Helicobacter pylori IgM ELISA

Helicobacter pylori IgM ELISA Instructions for Use Helicobacter pylori IgM ELISA Enzyme immunoassays for the qualitative and quantitative determination of IgM antibodies against Helicobacter pylori in human serum and plasma. RE56391

Mehr

25-OH-Vitamin D ELISA

25-OH-Vitamin D ELISA Arbeitsanleitung 25-OH-Vitamin D ELISA Enzymimmunoassay zur quantiativen Bestimmung von total 25-OH-Vitamin D (Vitamin D2 und Vitamin D3) in humanem Serum. RE53041 96 2-8 C I B L I N T E R N A T I O N

Mehr

Respiratory syncytial virus IgM ELISA

Respiratory syncytial virus IgM ELISA Arbeitsanleitung Respiratory syncytial virus IgM ELISA Enzymimmunoassays (Mikrotiterstreifen) zur qualitativen und quantitativen Bestimmung von IgM Antikörpern gegen Respiratory Syncytial Virus (RSV) in

Mehr

Aldosterone ELISA. Enzyme immunoassay for the quantitative determination of Aldosterone in human serum, plasma and urine.

Aldosterone ELISA. Enzyme immunoassay for the quantitative determination of Aldosterone in human serum, plasma and urine. Arbeitsanleitung Aldosterone ELISA Enzyme immunoassay for the quantitative determination of Aldosterone in human serum, plasma and urine. RE52301 96 2-8 C I B L I N T E R N A T I O N A L G M B H Flughafenstrasse

Mehr

Adenovirus IgA ELISA

Adenovirus IgA ELISA Arbeitsanleitung Adenovirus IgA ELISA Enzymimmunoassays (Mikrotiterstreifen) zur qualitativen und quantitativen Bestimmung von IgA Antikörpern gegen Adenovirus in humanem Serum und Plasma. RE56561 12x8

Mehr

Mycoplasma pneumoniae IgA ELISA

Mycoplasma pneumoniae IgA ELISA Arbeitsanleitung Mycoplasma pneumoniae IgA ELISA Enzymimmunoassay (Mikrotiterstreifen) zur qualitativen und quantitativen Bestimmung von IgA Antikörpern gegen Mycoplasma pneumoniae in humanem Serum und

Mehr

QuickGEN Sample Preparation Filtration

QuickGEN Sample Preparation Filtration QuickGEN Filtration INSTITUT FÜR ANGEWANDTE LABORANALYSEN GMBH QuickGEN Sample Preparation Filtration Art. Nr./ Order No.: FSE 0100 Version 11/16 GEN-IAL GmbH Tel: 0049 2241 2522980 Fax: 0049 2241 2522989

Mehr

CCP-Ab ELISA RE C. Arbeitsanleitung

CCP-Ab ELISA RE C. Arbeitsanleitung Arbeitsanleitung CCP-Ab ELISA Enzymimmunoassay für die quantitative Bestimmung von IgG Autoantikörpern gegen zyklische citrullinierte Peptide (CCP) in humanem Serum oder Plasma. RE75791 96 2-8 C I B L

Mehr

RIDASCREEN Foodscreen Blood Collection Card Art. No.: A8025-BCC

RIDASCREEN Foodscreen Blood Collection Card Art. No.: A8025-BCC RIDASCREEN Foodscreen Blood Collection Card Art. No.: A8025-BCC Zubehör für RIDASCREEN Spec. IgG Foodscreen Accessory for RIDASCREEN Spec. IgG Foodscreen 1. filling in patient data Procedure: 2. filling

Mehr

anti Sperm Ab ELISA IBL GESELLSCHAFT FÜR IMMUNCHEMIE UND IMMUNBIOLOGIE MBH FLUGHAFENSTRASSE 52a D-22335 HAMBURG GERMANY

anti Sperm Ab ELISA IBL GESELLSCHAFT FÜR IMMUNCHEMIE UND IMMUNBIOLOGIE MBH FLUGHAFENSTRASSE 52a D-22335 HAMBURG GERMANY Instructions for Use anti Sperm Ab ELISA Enzyme immunoassay for the in-vitro-diagnostic quantitative determination of autoantibodies against surface antigens of spermatozoa in human serum. RE52029 96 2-8

Mehr

Rubella virus IgG ELISA

Rubella virus IgG ELISA Arbeitsanleitung Rubella virus IgG ELISA Enzymimmunoassay für die quantitative Bestimmung von IgG Antikörpern gegen das Röteln Virus in humanem Serum und Plasma. RE58341 96 2-8 C I B L I N T E R N A T

Mehr

Epstein-Barr virus EA IgG ELISA

Epstein-Barr virus EA IgG ELISA Arbeitsanleitung Epstein-Barr virus EA IgG ELISA Enzymimmunoassays (Mikrotiterstreifen) für die qualitative und quantitative Bestimmung von IgG Antikörpern gegen das early antigen (EA) des Epstein-Barr

Mehr

Lipoprotein (a) ELISA

Lipoprotein (a) ELISA Arbeitsanleitung Lipoprotein (a) ELISA Enzymimmunoassay für die quantitative Bestimmung von Lipoprotein a [Lp(a)] in humanem Serum, Citrat und EDTA Plasma. RE59011 96 2-8 C I B L I N T E R N A T I O N

Mehr

Mycobacterium tuberculosis IgG ELISA

Mycobacterium tuberculosis IgG ELISA Arbeitsanleitung Mycobacterium tuberculosis IgG ELISA Enzymimmunoassay zur qualitativen und quantitativen Bestimmung von IgG Antikörpern gegen Mycobacterium tuberculosis in humanem Serum und Plasma. RE56591

Mehr

Serotonin high sensitive ELISA

Serotonin high sensitive ELISA Arbeitsanleitung Serotonin high sensitive ELISA Hochsensitiver Enzymimmunoassay für die quantitative Bestimmung von Serotonin in niedrig konzentrierten Proben und für kleine Probenvolumina. RE59141 96

Mehr

Testanleitung. Quantitative Bestimmung der Tumor M2-PK Stuhltest Instruction Manual. Quantitative Determination of the Tumor M2-PK Stool Test

Testanleitung. Quantitative Bestimmung der Tumor M2-PK Stuhltest Instruction Manual. Quantitative Determination of the Tumor M2-PK Stool Test Testanleitung Quantitative Bestimmung der Tumor M2-PK Stuhltest Instruction Manual Quantitative Determination of the Tumor M2-PK Stool Test Tumor M2-PK ELISA Stuhltest/Stool Test REF Artikel-Nr./Catalog

Mehr

17beta-Estradiol Luminescence Immunoassay

17beta-Estradiol Luminescence Immunoassay Instructions for Use 17beta-Estradiol Luminescence Immunoassay Luminescence immunoassay for the in-vitro diagnostic quantitative determination of free 17beta-estradiol in human saliva. RE62041 96 2-8 C

Mehr

Toxoplasma gondii IgM ELISA

Toxoplasma gondii IgM ELISA Arbeitsanleitung Toxoplasma gondii IgM ELISA Enzymimmunoassay für die quantitative Bestimmung von IgM Antikörpern gegen Toxoplasma gondii in humanem Serum und Plasma. RE58381 96 2-8 C I B L I N T E R N

Mehr

Progesterone Saliva ELISA

Progesterone Saliva ELISA Arbeitsanleitung Progesterone Saliva ELISA Enzymimmunoassay zur quantitativen Bestimmung von Progesteron in humanem Speichel. RE52281 96 2-8 C I B L I N T E R N A T I O N A L G M B H Flughafenstrasse 52a

Mehr

Parainfluenza 1/2/3 IgA ELISA

Parainfluenza 1/2/3 IgA ELISA Instructions for Use Parainfluenza 1/2/3 IgA ELISA Enzyme immunoassays (microtiter strips) for the qualitative and quantitative determination of IgA antibodies against Parainfluenza virus 1,2 and 3 in

Mehr

alpha-amylase Saliva Assay

alpha-amylase Saliva Assay Arbeitsanleitung alpha-amylase Saliva Assay Enzymatischer Assay zur Bestimmung der Alpha-Amylase-Aktivität in humanem Speichel. RE80111 96 2-8 C I B L I N T E R N A T I O N A L G M B H Flughafenstrasse

Mehr

Simplex Easy DNA kit

Simplex Easy DNA kit Simplex Easy DNA INSTITUT FÜR ANGEWANDTE LABORANALYSEN GMBH Simplex Easy DNA kit Schnelle DNA-Extraktion aus Bakterien und Hefen Fast DNA-extraction from bacteria and yeast Art. Nr./ Order No.: SE 0100;

Mehr

Estrone ELISA. Enzymimmunoassay für die direkte quantitative Bestimmung von Estron in humanem Serum.

Estrone ELISA. Enzymimmunoassay für die direkte quantitative Bestimmung von Estron in humanem Serum. Arbeitsanleitung Estrone ELISA Enzymimmunoassay für die direkte quantitative Bestimmung von Estron in humanem Serum. DB52051 96 28 C I B L I N T E R N A T I O N A L G M B H Flughafenstrasse 52a Phone:

Mehr

BCA Protein Assay (Pierce)

BCA Protein Assay (Pierce) BCA Protein Assay (Pierce) BCA assay: geeignet für geringe Proteinmengen!!! Die Proteine bilden mit Cu2+-Ionen in alkalischer Lösung einen Komplex (Biuret-Reaktion). Die Cu2+-Ionen des Komplexes werden

Mehr

Herpes simplex virus 1/2 IgG ELISA

Herpes simplex virus 1/2 IgG ELISA Arbeitsanleitung Herpes simplex virus 1/2 IgG ELISA Enzymimmunoassays (Mikrotiterstreifen) zur qualitativen und quantitativen Bestimmung von IgG Antikörpern gegen Herpes simplex Virus 1 und 2 in humanem

Mehr

Hantavirus Hantaan IgG/IgM ELISA

Hantavirus Hantaan IgG/IgM ELISA Arbeitsanleitung Hantavirus Hantaan IgG/IgM ELISA Enzymimmunoassay zur qualitativen Bestimmung von Hantavirus IgG-/IgM-Antikörpern gegen den Hantaan-Serotyp des Hantavirus in humanem Serum. RE57451 96

Mehr

5alpha- Dihydrotestosterone ELISA

5alpha- Dihydrotestosterone ELISA Arbeitsanleitung 5alpha Dihydrotestosterone ELISA Enzymimmunoassay zur quantitativen Bestimmung von 5alpha Dihydrotestosteron in humanem Serum. DB52021 96 28 C I B L I N T E R N A T I O N A L G M B H Flughafenstrasse

Mehr

Measles IgM µ-capture ELISA

Measles IgM µ-capture ELISA Instructions for Use Measles IgM µ-capture ELISA Enzyme immunoassay for the qualitative and quantitative determination of IgM antibodies to measles virus in human serum and plasma RE57151 96 2-8 C IBL

Mehr

Candida albicans IgM ELISA

Candida albicans IgM ELISA Arbeitsanleitung Candida albicans IgM ELISA Enzymimmunoassays zur qualitativen und quantitativen Bestimmung von IgM Antikörpern gegen Candida ablicans in humanem Serum und Plasma. RE56181 12x8 2-8 C I

Mehr

CRP ELISA. Enzyme immunoassay for the in-vitro-diagnostic quantitative determination of C-reactive protein in human serum and plasma.

CRP ELISA. Enzyme immunoassay for the in-vitro-diagnostic quantitative determination of C-reactive protein in human serum and plasma. Instructions for Use CRP ELISA Enzyme immunoassay for the in-vitro-diagnostic quantitative determination of C-reactive protein in human serum and plasma. EU59131 96 2-8 C I B L I N T E R N A T I O N A

Mehr

N-MID Osteocalcin ELISA

N-MID Osteocalcin ELISA Instructions for Use N-MID Osteocalcin ELISA OD53121 For illustrative purposes only. To perform the assay the instructions for use provided with the kit have to be used. Distributed by: I B L I N T E R

Mehr

Epstein-Barr virus EBNA-1 IgA ELISA

Epstein-Barr virus EBNA-1 IgA ELISA Arbeitsanleitung Epstein-Barr virus EBNA-1 IgA ELISA Enzymimmunoassays (Mikrotiterstreifen) zur qualitativen und quantitativen Bestimmung von IgA Antikörpern gegen das Epstein-Barr Virus nuclear antigen

Mehr

Rheumatoid Factor IgA ELISA

Rheumatoid Factor IgA ELISA Arbeitsanleitung Rheumatoid Factor IgA ELISA Enzymimmunoassay zur quantitativen Bestimmung von IgA Rheumafaktoren (RF) in humanem Serum oder Plasma. RE75761 96 Nur zur allgemeinen Veranschaulichung. Den

Mehr

Mycoplasma pneumoniae IgM ELISA

Mycoplasma pneumoniae IgM ELISA Arbeitsanleitung Mycoplasma pneumoniae IgM ELISA Enzymimmunoassays (Mikrotiterstreifen) zur qualitativen und quantitativen Bestimmung von IgM Antikörpern gegen Mycoplasma pneumoniae in humanem Serum und

Mehr

DHEA Luminescence Immunoassay

DHEA Luminescence Immunoassay Arbeitsanleitung DHEA Luminescence Immunoassay Lumineszenz Immunoassay zur quantitativen Bestimmung von aktivem freiem DHEA in humanem Saliva und Gesamt-DHEA in verdünntem humanem Serum. RE62051 RE62059

Mehr

Rat Control HZ-DEV99RC 001 2016-04. Rat control sera suitable for internal control quality of the following HOELZEL ELISAs:

Rat Control HZ-DEV99RC 001 2016-04. Rat control sera suitable for internal control quality of the following HOELZEL ELISAs: Rat Control Rat control sera suitable for internal control quality of the following HOELZEL ELISAs: Testosterone rat/mouse Corticosterone rat/mouse Prolactin rat TSH rat (HZ-DEV9911) (HZ-DEV9922) (HZ-DEV9966)

Mehr

IRT neonatal screening ELISA

IRT neonatal screening ELISA Arbeitsanleitung IRT neonatal screening ELISA Enzymimmunoassay zur quantitativen Bestimmung von immunreaktivem Trypsin (IRT) in Trockenblutproben von humanen Neugeborenen. Für das Neugeborenenscreening

Mehr

17-OH-Progesterone ELISA

17-OH-Progesterone ELISA Arbeitsanleitung 17-OH-Progesterone ELISA Enzymimmunoassay zur quantitativen in-vitro-bestimmung von 17-OH-Progesteron in humanem Serum und Plasma. RE52071 96 2-8 C I B L I N T E R N A T I O N A L G M

Mehr

dsdna-ab RIA Radioimmunoassay zur quantitativen Bestimmung von Autoantikörpern gegen dsdna in humanem Serum.

dsdna-ab RIA Radioimmunoassay zur quantitativen Bestimmung von Autoantikörpern gegen dsdna in humanem Serum. Arbeitsanleitung dsdna-ab RIA Radioimmunoassay zur quantitativen Bestimmung von Autoantikörpern gegen dsdna in humanem Serum. RE19011 100 2-8 C I B L I N T E R N A T I O N A L G M B H Flughafenstrasse

Mehr

Diphtheria-Ab ELISA. Enzymimmunoassay zur quantitativen Bestimmung von Diphtherie-Antikörpern in humanem Serum und Plasma.

Diphtheria-Ab ELISA. Enzymimmunoassay zur quantitativen Bestimmung von Diphtherie-Antikörpern in humanem Serum und Plasma. Arbeitsanleitung Diphtheria-Ab ELISA Enzymimmunoassay zur quantitativen Bestimmung von Diphtherie-Antikörpern in humanem Serum und Plasma. RE57431 96 2-8 C I B L I N T E R N A T I O N A L G M B H Flughafenstrasse

Mehr

CRP ELISA. Enzymimmunassay zur quantitativen Bestimmung von C-reaktivem Protein in humanem Serum und Plasma.

CRP ELISA. Enzymimmunassay zur quantitativen Bestimmung von C-reaktivem Protein in humanem Serum und Plasma. Arbeitsanleitung CRP ELISA Enzymimmunassay zur quantitativen Bestimmung von C-reaktivem Protein in humanem Serum und Plasma. EU59131 96 2-8 C I B L I N T E R N A T I O N A L G M B H Flughafenstrasse 52a

Mehr

Rheumatoid Factor IgM ELISA

Rheumatoid Factor IgM ELISA Arbeitsanleitung Rheumatoid Factor IgM ELISA Enzymimmunoassay zur quantitativen Bestimmung von IgM Rheumafaktoren (RF) in humanem Serum oder Plasma. RE75781 96 Nur zur allgemeinen Veranschaulichung. Den

Mehr

Parainfluenza 1/2/3 IgA ELISA

Parainfluenza 1/2/3 IgA ELISA Arbeitsanleitung Parainfluenza 1/2/3 IgA ELISA Enzymimmunoassays (Mikrotiterstreifen) zur qualitativen und quantitativen Bestimmung von IgA Antikörpern gegen Parainfluenza Virus 1, 2 und 3 in humanem Serum

Mehr

FT3 ELISA. Enzymimmunoassay für die direkte quantitative Bestimmung von freiem Trijodthyronin (ft3) in Humanserum oder Plasma.

FT3 ELISA. Enzymimmunoassay für die direkte quantitative Bestimmung von freiem Trijodthyronin (ft3) in Humanserum oder Plasma. Arbeitsanleitung FT3 ELISA Enzymimmunoassay für die direkte quantitative Bestimmung von freiem Trijodthyronin (ft3) in Humanserum oder Plasma. RE55231 96 2-8 C I B L I N T E R N A T I O N A L G M B H Flughafenstrasse

Mehr

Candida albicans IgA ELISA

Candida albicans IgA ELISA Arbeitsanleitung Candida albicans IgA ELISA Enzymimmunoassay zur qualitativen und quantitativen Bestimmung von IgA Antikörpern gegen Candida ablicans in humanem Serum und Plasma. RE56161 12x8 2-8 C I B

Mehr

Estriol Saliva ELISA

Estriol Saliva ELISA Arbeitsanleitung Estriol Saliva ELISA Enzymimmunoassay für die quantitative Bestimmung von Estriol in humanem Speichel. RE52291 96 2-8 C I B L I N T E R N A T I O N A L G M B H Flughafenstrasse 52a Phone:

Mehr

Epstein-Barr virus VCA IgA ELISA

Epstein-Barr virus VCA IgA ELISA Arbeitsanleitung Epstein-Barr virus VCA IgA ELISA Enzymimmunoassays (Mikrotiterstreifen) zur qualitativen und quantitativen Bestimmung von IgA Antikörpern gegen das viral capsid antigen (VCA) des Epstein-Barr

Mehr

Aquaporin-4 (AQP4) Ab ELISA

Aquaporin-4 (AQP4) Ab ELISA Arbeitsanleitung Aquaporin-4 (AQP4) Ab ELISA Enzymimmunoassay für die quantitative Bestimmung von AQP4-Autoantikörpern in Serum REF EA111/96 12 x 8 I V D 2 8 C DLD Gesellschaft für Diagnostika und medizinische

Mehr

Noradrenalin ELISA RE C. Arbeitsanleitung

Noradrenalin ELISA RE C. Arbeitsanleitung Arbeitsanleitung Noradrenalin ELISA Manueller und automatischer Enzymimmunoassay für die quantitative In-vitro-Bestimmung von Noradrenalin (Norepinephrin) in humanem Plasma und Urin. RE59261 96 2-8 C I

Mehr

TBE virus (FSME) IgG ELISA

TBE virus (FSME) IgG ELISA Arbeitsanleitung TBE virus (FSME) IgG ELISA Enzymimmunoassay zur Bestimmung von IgG-Antikörpern gegen das FSME-Virus in humanem Serum und Plasma. RE57401 96 2-8 C I B L I N T E R N A T I O N A L G M B

Mehr